Ahiba IR User's Guide
Equipment Controls and Indicators
•
15
2.
Tighten the dog point socket head screws using the hex screwdriver
provided with the unit.
CAUTION
Do not replace hex screwdriver with a hex key.
This will cause the lid to be over tightened.
3.
Install the beakers on the wheel.
Use the following guidelines when installing the beakers on the wheel:
•
Do not mix beaker sizes in a single process. All beakers loaded on the
wheel must be the same size.
•
The Ahiba IR uses custom designed pressure-tested beakers. The
seal between the beaker and the lid is accomplished with an o-ring
seal. This o-ring is a maintenance part and additional o-rings are
supplied with the unit.
•
Over tightening of the beaker will damage the o-ring and cause damage
to the lid. If additional tightening is required to ensure the seal, it is
time to replace the o-rings.
See also Appendix, System Maintenance
for the maintenance schedule.