Water Flow Meter Calibration:
A. Ensure tube is filled with water.
Go to:
MANUAL MODE, MANUAL FUNCTION, FILL CW 50ml
to fill tube with
water.
B. Drain beaker.
Go to:
MANUAL MODE, MANUAL FUNCTION, DRAIN XXsec.
Ensure
the beaker lid is on to build pressure up so water is pushed out of the
beaker.
C. Release pressure in the chamber by pressing the manual “Pressure Relief” valve
switch located next to the chamber as shown below.
Note: It is important to maintain atmospheric pressure during calibration for
accuracy.
ahiba
49
AHIBA MULTIPRECISE TC - Calibration (Water Flow Meter)
Содержание Ahiba Ahiba MULTIPRECISE TC
Страница 1: ......
Страница 3: ......
Страница 6: ...ahiba 5 Safety Precautions Safety Precautions and W and Warnings arnings ...
Страница 7: ......
Страница 11: ......
Страница 12: ...ahiba 11 Introduction T Introduction Typical ypical Applications Applications ...
Страница 13: ......
Страница 16: ...ahiba 15 Main Components Main Components ...
Страница 17: ......
Страница 25: ......
Страница 26: ...ahiba 25 Installation and S Installation and Start Up tart Up ...
Страница 27: ......
Страница 35: ......
Страница 36: ...ahiba 35 T Technical Data echnical Data ...
Страница 37: ......
Страница 41: ......
Страница 42: ...ahiba 41 Procedure for Dosing Procedure for Dosing ...
Страница 43: ......
Страница 48: ...ahiba 47 Calibration Calibration ...
Страница 49: ......
Страница 53: ......
Страница 54: ...ahiba 53 Controller Controller ...
Страница 55: ......
Страница 81: ......
Страница 82: ...ahiba 81 Protection Protection ...
Страница 83: ......
Страница 85: ......
Страница 86: ...ahiba 85 Maintenance Maintenance ...
Страница 87: ......
Страница 93: ......
Страница 94: ...ahiba 93 Appendices Appendices ...
Страница 95: ......