Colibrick A/D Converter
4 Using the Colibrick
4.2 Method Setup
dialogs depend on the type of analysis (measuring
conditions), so they can be specific for various types of analysis. This is
the reason for them to be a part of the method.
Caution:
The parameters can't be changed during analysis.
4.2.1 Method Setup - Acquisition
Fig 14: Method Setup - Acquisition (Colibrick)
Select Detector
Selects the detector whose parameters will be set. If there is only one
detector (channel) configured on the
Clarity
Instrument, this control will
not be accessible.
Range
Selects the input voltage range (in mV). The selected range corresponds
to the maximum output voltage of the detector that the A/D converter can
still process.
Available ranges for
Colibrick
converter are
156
,
1250
and
12123 mV
.
Sample Rate
. The available sampling rates are
6.3
,
12.5
,
25
,
50
,
100
,
200
and
400
samples per second (for 50Hz supply
- 17 -