Appendix A
86
shows the frequency response using a sampling rate of 4800 Samples/s.
Figure 19: Frequency Response of the DT9824 When Using a Sampling Rate of 4800 Samples/s
As you can see, the frequency response of the DT9824 gently rolls off from DC due to the
filtering of the Sigma-Delta A/D converter. You can use software to apply a correction factor
to the acquired data to compensate for the filter response of the A/D converter. The correction
factor depends on the input frequency of the signal you are trying to measure and the
sampling rate of the DT9824.
shows the gain correction factors for typical input
frequencies and sampling rates.
–3 dB point = 1104 Hz
Содержание DT9824
Страница 1: ...DT9824 UM 23782 G User s Manual Title Page ...
Страница 4: ......
Страница 8: ...Contents 8 ...
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Features 14 Supported Software 16 Getting Started Procedure 18 ...
Страница 19: ...Part 1 Getting Started ...
Страница 20: ......
Страница 21: ...21 2 Installing the Module Unpacking 23 Attaching Modules to the Computer 24 Configuring the Driver 27 ...
Страница 28: ...Chapter 2 28 ...
Страница 44: ...Chapter 4 44 ...
Страница 45: ...45 Part 2 Using Your Module ...
Страница 46: ...Chapter 46 ...
Страница 47: ...47 5 Principles of Operation Analog Input Features 49 Digital I O Features 54 ...
Страница 48: ...Chapter 5 48 Figure 13 shows a block diagram of the DT9824 module Figure 13 Block Diagram of the DT9824 Module ...
Страница 56: ...Chapter 5 56 ...
Страница 68: ...Chapter 6 68 ...
Страница 69: ...69 7 Calibration Overview 70 Using the Auto Calibration Procedure 71 Using the Manual Calibration Procedure 72 ...
Страница 74: ...Chapter 7 74 ...
Страница 75: ...75 8 Troubleshooting General Checklist 76 Technical Support 78 If Your Module Needs Factory Service 79 ...
Страница 80: ...Chapter 8 80 ...
Страница 92: ...Appendix A 92 ...
Страница 98: ...Appendix B 98 ...