Calibration
73
d.
Once the Running Average value has settled, adjust the gain by entering values in the
Gain edit box, or by clicking the up/down buttons until the Running Average value
matches the value you applied.
e.
Repeat steps 4a to 4d, selecting the next gain in the list, until the gain values are
calibrated for each gain setting.
5.
Repeat steps 2 to 4 for the remaining frequency until the offset and gain values are
calibrated for each frequency for the selected channel.
6.
Repeat steps 1 to 5 for the remaining A/D channels that you want to calibrate.
Note:
At any time, you can click
Restore Factory Settings
to reset the A/D calibration
values to their original factory settings. This process will undo any auto or manual calibration
settings.
Содержание DT9824
Страница 1: ...DT9824 UM 23782 G User s Manual Title Page ...
Страница 4: ......
Страница 8: ...Contents 8 ...
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Features 14 Supported Software 16 Getting Started Procedure 18 ...
Страница 19: ...Part 1 Getting Started ...
Страница 20: ......
Страница 21: ...21 2 Installing the Module Unpacking 23 Attaching Modules to the Computer 24 Configuring the Driver 27 ...
Страница 28: ...Chapter 2 28 ...
Страница 44: ...Chapter 4 44 ...
Страница 45: ...45 Part 2 Using Your Module ...
Страница 46: ...Chapter 46 ...
Страница 47: ...47 5 Principles of Operation Analog Input Features 49 Digital I O Features 54 ...
Страница 48: ...Chapter 5 48 Figure 13 shows a block diagram of the DT9824 module Figure 13 Block Diagram of the DT9824 Module ...
Страница 56: ...Chapter 5 56 ...
Страница 68: ...Chapter 6 68 ...
Страница 69: ...69 7 Calibration Overview 70 Using the Auto Calibration Procedure 71 Using the Manual Calibration Procedure 72 ...
Страница 74: ...Chapter 7 74 ...
Страница 75: ...75 8 Troubleshooting General Checklist 76 Technical Support 78 If Your Module Needs Factory Service 79 ...
Страница 80: ...Chapter 8 80 ...
Страница 92: ...Appendix A 92 ...
Страница 98: ...Appendix B 98 ...