Appendix A
84
Noise
shows the maximum noise measurement at 10 Hz with a gain of 1.
Figure 17: Maximum Noise Measurement During Typical Usage at 10 Hz with a Gain of 1
shows the typical RMS noise, in microvolts, of the DT9824 at sampling rates of
10 Samples/s, 100 Samples/s, 960 Samples/s, and 4800 Samples/s and gains of 1, 8, 16, and
32.
shows this data in graphical form.
Table 23: DT9824 System Noise
Gain
10 Samples/s
100 Samples/s
960 Samples/s
4800 Samples/s
1
3.6
μ
Vrms
12.0
μ
Vrms
25.0
μ
Vrms
63.0
μ
Vrms
8
0.7
μ
Vrms
2.0
μ
Vrms
4.2
μ
Vrms
11.0
μ
Vrms
16
0.6
μ
Vrms
1.8
μ
Vrms
2.8
μ
Vrms
7.5
μ
Vrms
32
0.4
μ
Vrms 1.3
μ
Vrms
2.3
μ
Vrms
6.5
μ
Vrms
Содержание DT9824
Страница 1: ...DT9824 UM 23782 G User s Manual Title Page ...
Страница 4: ......
Страница 8: ...Contents 8 ...
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Features 14 Supported Software 16 Getting Started Procedure 18 ...
Страница 19: ...Part 1 Getting Started ...
Страница 20: ......
Страница 21: ...21 2 Installing the Module Unpacking 23 Attaching Modules to the Computer 24 Configuring the Driver 27 ...
Страница 28: ...Chapter 2 28 ...
Страница 44: ...Chapter 4 44 ...
Страница 45: ...45 Part 2 Using Your Module ...
Страница 46: ...Chapter 46 ...
Страница 47: ...47 5 Principles of Operation Analog Input Features 49 Digital I O Features 54 ...
Страница 48: ...Chapter 5 48 Figure 13 shows a block diagram of the DT9824 module Figure 13 Block Diagram of the DT9824 Module ...
Страница 56: ...Chapter 5 56 ...
Страница 68: ...Chapter 6 68 ...
Страница 69: ...69 7 Calibration Overview 70 Using the Auto Calibration Procedure 71 Using the Manual Calibration Procedure 72 ...
Страница 74: ...Chapter 7 74 ...
Страница 75: ...75 8 Troubleshooting General Checklist 76 Technical Support 78 If Your Module Needs Factory Service 79 ...
Страница 80: ...Chapter 8 80 ...
Страница 92: ...Appendix A 92 ...
Страница 98: ...Appendix B 98 ...