10. TECHNICAL SPECIFICATION
Information
Display Mode
Pulse Oxygen Saturation
(
SpO2
)
2-digit digital - OLED display
Pulse Rate
(
PR
)
3-digit digital - OLED display
Pulse Intensity (bar-graph)
bar-graph - OLED display
SpO2 Parameter Specification
Measuring range
0%
~
100%, (resolution: 1%)
Accuracy
70%
~
100%
:
±2%, below 70% unspecified
Average value
Calculate the Average value in every 4 measure value. The deviation
between average value and true value does not exceed 1%.
Pulse Parameter Specification
Measuring range
30bpm
~
250bpm, (the resolution is 1bpm)
Accuracy
±2bpm or±2% (select larger)
Average pulse
rate
Calculates the average pulse rate every 4th cardio cycle.
Deviation between average value and true value does not exceed 1%.
Safety Type
Interior battery, BF type
Pulse Intensity
Range
Continuous bar-graph display.
Battery Requirement
2 x AA
Oximeter Probe
Wavelength:660nm 880nm
Dimensions and Weight
Dimensions
110(l) × 60(w) × 23(h) mm
Weight
About 180g (with batteries)
Appendix
State
Alarm condition delay
Alarm signal generation delay
Low voltage alarm
1s
20ms
Spo2 alarm
330ms
20ms
Pulse rate alarm
330ms
20ms
Probe error alarm
16ms
20ms