14
Menu operations:
On the measuring interface, press the “menu
button “can enter the menu of figure 9.Users can
adjust the settings through the main menu, such as
alarm, pulse sound indication, backlight, data
storage, data transmission (with the use of data
line), the specific method is as follows:
Alarm setting
On the main menu interface, press the “up button
"or" down button" to select "Alarm”, then press the
“left button" or “right button “to enter the alarm
setting menu of figure 10:
The highest/lowest alarm limit setting
Press the “up button "or" down button "to choose
the parameter to be adjusted, then press the "left
button” or “right button" to change data. Each
press of the "left button" or "right button", the
data will raise or descend for one time
accordingly.
If the alarm function is on, the device will provide medium-priority alarm signal when the
data of SpO2 or pulse rate is beyond the limit. Intermittent alarm will occur and the
measurement shows in yellow.
Medium priority indicating that prompt operator response is required.
The alarm state setting
Press the “up button" or "down button" to select "Alarm", then choose the alarm state
(on/off) by pressing the "left button" or "right button", choose “on” to turn on the alarms,
and choose “off” to turn off the alarms for good.
Exit the Alarm settings
Press the "menu button" to exit the Alarm Settings Menu.
Fig.9 Main Menu Interface
Fig.10 Alarm Setting Menu