8. Maintenance
The proposed maintenance is only a suggestion. Depending on the use of the product and the
operating environment, this may need to be revised more often. Daray would strongly recommend
against “quick fixes” with tape, etc. If you cannot resolve a problem then please contact our
helpline on
0800 878 9864
or email
8.1 Maintenance schedule
Weekly checks
should include:-
Check for any Mechanical wear
Check all outer cables and accessories are in good condition
Monthly checks to
include:-
Check the Internal balloon of any reusable blood pressure cuffs.
Check if all the monitoring functions of the monitor can work normally
so as to make sure that the monitor is in proper working condition
Twelve monthly
checks to
include:-
Electrical Safety Certification
NIBP recalibration every 2 years - This product is calibrated before
leaving the factory, recalibration of the NIBP measurement should be
conducted every two years (or as frequently as dictated by your Hospital
Procedures Policy)
In case of any damage, abnormal function, potential safety danger or exception, do not use the
device on a patient. Contact the technician in your hospital or the supplier immediately.
Make sure the qualified service personnel have implemented a complete inspection, including any
functional safety checks, after the monitor has been used for 6-12 consecutive months, or after
monitor servicing or a computer system upgrades. This is to ensure the normal operation of the
V460.
8.2 Safety Precautions
Failure on the part of the responsible party using the V460; to implement a satisfactory
maintenance schedule may cause undue equipment failure and possible health hazards.
Warning: Any safety inspections or maintenance that require access to the internal aspects of
the monitor, must be performed by trained and authorized personnel only. Otherwise,
equipment failure and possible health hazards may be caused, please contact Daray for more
details.
If the equipment is scratched or dented this should not impact the usability of our products
however we advise that you do not attempt to cover over any damage as this will impact upon the
effectiveness of any cleaning you do.
Simply dampen a cloth with an alcohol based liquid when cleaning and pay special attention to
any damaged areas.
Do not use the V460 if there is any visible damage or cracks in the monitor casing
or its accessories
Do not use any sensor or cable that may be damaged or that has deteriorated
Содержание V460S Series
Страница 1: ...V460S Series Ambulatory NIBP Patient Monitor with SpO2 User Manual Installation Guide QAM V460S 0915 4...
Страница 2: ...Page 1 of 49...
Страница 10: ...Page 9 of 49 2 General Operation Guide Section...
Страница 15: ...Page 14 of 49 4 NIBP Operation...
Страница 20: ...Page 19 of 49 5 SpO2 Operation...
Страница 24: ...Page 23 of 49 6 Software Operation Initial Set up Guide Section...
Страница 38: ...Page 37 of 49 7 First Time Setup and the on board menu functions...