Страница 3: ...lling the Ceiling plate 17 4 3 4 Wiring and transformer layout 19 4 3 5 Primary rotation arm installation 20 4 3 6 Suspension arm Installation 22 4 3 7 Attaching the Ceiling Cover 26 4 4 Wall Mounted...

Страница 4: ...evice and all of its features appropriately It also contains all the information you need for the installation and maintenance of the light All versions are delivered in special to type packaging whic...

Страница 5: ...ll 0333 321 0971 This product was designed manufactured in Great Britain by Daray Ltd Edison House Robian Way Swadlincote Derbyshire DE11 9DH 1 1 Supported Lights The Following Models are supported in...

Страница 6: ...nergy with wavelength less than 400mm w m 0 002 Input Rating 220V 50Hz Power 70W Average Working Life 50 000 hours Warranty Period 3 Years NHS E Class Code FBU FKR Commodity Code 9018 908400 GMDN Code...

Страница 7: ...lowing items in the specified quantities SL730 Mobile Mounted Pack List Qty Mobile Base with 4 wheels 2 lockable with 5m power cable 1 Lower Upstand tubing 1 Upper Upstand tubing joint spigot 1 Suspen...

Страница 8: ...sion Arm Upper Vertical Tubing Spigot Lower Vertical Tubing Mobile Base Lighthead 3 Operation 3 1 Range of Motion The range of motion between the SL730 series varies This is an image depicting the ran...

Страница 9: ...vice uninterrupted for up to 10 hours 4 Disconnect the electrical supply and re attach the battery to the light allow the light while disconnected from its primary power source to run on the battery u...

Страница 10: ...and the minus button to lower the intensity dimmer Similarly for the Colour control use the plus button to increase the Colour Temperature and the minus button to lower the Colour Temperature range b...

Страница 11: ...lt a qualified electrician Live Electric Circuitry All personnel working around live circuitry should be qualified and following proper procedures Environmental Responsibility Once our products are in...

Страница 12: ...ur kit you should identify two pieces of the upstand Feed the wire see fig 3 from the mobile base through the shorter of the two pieces noting that the end with the four screw holes should be uppermos...

Страница 13: ...stand to face the front of the base see fig 6 and then secure at the base using the three allen screws provided These need to be tight Fig 7 Locate the Suspension arm see fig 7 then make the electrica...

Страница 14: ...o remove the small grub screw in order to fit these two parts together see fig 11 If you have removed this part it is important that this part be refitted after assembly see fig 13 Fig 11 Secure toget...

Страница 15: ...e 8 for instructions Check that all lights illuminate see fig 15 Fig 15 Check the arm and head for smooth and correct movement If the head is drifting or if the movement of the arm is stiff use a hex...

Страница 16: ...height Live Electric Circuitry All personnel working around live circuitry should be qualified and following proper procedures Heavy or Difficult fittings Fitting the ceiling plate can be difficult a...

Страница 17: ...that the uni strut is cross braced to ensure it is secure and does not move Welded to structure The fixing plate is the component that would be welded to the structure any welding must be assessed to...

Страница 18: ...h to the ceiling super structure instead of the ceiling plate Whichever fixing method used it is highly advisable that the plate being fixed to the ceiling is secured with a minimum of 4 bolting point...

Страница 19: ...the vertical plane This can be achieved by adjusting the 2nd level of nuts the adjusting nut Once this adjustment has been done tighten the 1st level of nuts down to the ceiling plate to lock the plat...

Страница 20: ...al movement parts of the switch should meet IEC447 requirements and the symbols should meet IEC60601 1 requirements Ideally the cable used to power the light should be of a 1 5mm twin and earth type D...

Страница 21: ...h cannot be modified after installation If you require a customised length of downtube please contact DARAY The primary rotation arm is fitted to a 360 degree arm boss system the arm boss is the secti...

Страница 22: ...ullet connectors coming from the transformer Fig 5 Note Before progressing onto fixing any other arms ensure that the rotation arm is attached securely Warning please observe the Cable Polarities and...

Страница 23: ...rm rotation arm we have just installed and second arm suspension arm observing polarity Fig 6 Fig 7 Making the electrical connection between the Rotation arm and the Suspension arm Fig 8 Resting the S...

Страница 24: ...these parts should arrive already connected to each other Make the electrical connection between the Hook arm and the Suspension arm attached to the upstand see fig 10 Warning without the lighthead a...

Страница 25: ...fig 13 Fig 11 Secure together with the three screws supplied see fig 12 Fig 12 Fig 13 Slide the blue rubber collar over this joint to finish see fig 14 Fig 14 The light can now be tested Switch on th...

Страница 26: ...ons detailed in maintenance section of this manual Power the light on using the button located on the light head Check that all lights illuminate Your new DARAY light is now ready for use To gain full...

Страница 27: ...ceiling mount and secure them together using the screws provided Once the two halves of the primary cover are connected work the cover over the cabling and then attach the secondary securing ring aro...

Страница 28: ...overning regulations The normal life span of the SL730 series is up to ten years We recommend the equipment should be discarded and recycled when reaching the end of its life These instructions are ba...

Страница 29: ...bility of the surface to be fixed to 4 4 3 Installing the wall plate Fig 2 Fig 3 To install the wall plate you may need more than 1 person Align the mount to the wall and mark your fixing points Ensur...

Страница 30: ...power to the mains connection and power the light on using the button located on the lighthead Check that all LED s illuminate The light can now be tested by applying power and checking the arms and h...

Страница 31: ...rm rotation arm we have just installed and second arm suspension arm observing polarity Fig 6 Fig 7 Making the electrical connection between the Rotation arm and the Suspension arm Fig 8 Resting the S...

Страница 32: ...these parts should arrive already connected to each other Make the electrical connection between the Hook arm and the Suspension arm attached to the upstand see fig 10 Warning without the lighthead a...

Страница 33: ...fig 13 Fig 11 Secure together with the three screws supplied see fig 12 Fig 12 Fig 13 Slide the blue rubber collar over this joint to finish see fig 14 Fig 14 The light can now be tested Switch on th...

Страница 34: ...ons detailed in maintenance section of this manual Power the light on using the button located on the light head Check that all lights illuminate Your new DARAY light is now ready for use To gain full...

Страница 35: ...that there is no visible cable wear All fixings to be checked and be of sound construction Check optional battery back ups for deformities and overheating 5 2 Safety Precautions For all cleaning work...

Страница 36: ...our BioProtect antimicrobial coating on painted finishes This is designed to assist with infection control and is not a replacement for cleaning CAUTION Our products are NOT waterproof The equipment...

Страница 37: ...The LED bulbs in the SL730 Series are high quality and should not need to be replaced in the lifetime of the light however if any bulbs need to be replaced please consult a qualified electrician or c...

Страница 38: ...b screw located on the opposite side of the arm boss to the main arm section Suspension arm all versions The suspension arms can be adjusted in several ways however its upward and downward movement is...

Страница 39: ...justed by using a flat blade tool in the grub screws located on each side of the arm boss Lighthead arm all versions The side turning movement of the lighthead arm is controlled by the central tension...

Страница 40: ...s in section 3 Warning without the lighthead attached the suspension arm will spring upwards rapidly Keep control of the arm at all times and avoid any obstructions in its path The SL730 is a complete...

Страница 41: ...are fully connected Poor Light output Power Button Please check the variable intensity is not set to its minimum Voltage in Verify that the correct mains voltage and correct transformer is being used...

Страница 42: ...ges may be applicable OTHER Any situation which is not covered by any of the above We will always try to help but we cannot normally offer a refund For additional clarification please refer to our ter...

Страница 43: ...Daray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from...

Страница 44: ...Page 43 of 43 Service Notes...

Страница 45: ...Page 44 of 43...

Страница 46: ...Page 45 of 43...

Страница 47: ...Page 46 of 43...

Страница 48: ...DARAY Ltd Edison House Robian Way Swadlincote Derbyshire United Kingdom DE11 9DH 44 0 800 878 9864 44 0 333 321 0973 sales daray co uk www daray co uk...

Отзывы: