To view memory contents, follow these steps:

Select Recall #0 button [4] and enter the number
of the test to be retrieved.

If you do not know the test number,
enter the last test number and scroll
through the list using Next [+] or
Previous [–] to find the test you want to review.

To print the result, press the quit button
[0], then press button [5]. A copy of the
result will be printed.

Note:

The result will be printed in the format

last selected through the Set function.

1 0

Vickers by Danfoss Target-Pro 2 Analyzer and Sampler Operation Manual V-PP-MC-0002-E

July 2023

Install software Pt. No. Target-Pro 2-W-30 on a suitable PC
running W indows 95

®

/NT 4.0 or better. Follow instructions

from the Target-Pro 2-View User Manual.

Results Download

1.

Connect RS232 cable to analyzer and serial port on PC.

2.

Switch the analyzer on.

3.

Switch PC on.

Launch Target-Pro 2-View software. From
the File menu, select Upload. Data
Transfer Screen will be displayed.
Transfer Data.

4.

Select log function from Analyzer Main/Test
Progress Screen [6].

5.

Press Transfer Log [1].

The analyzer will download all stored resultsfrom

memory to the software. After downloading, the
analyzer’s memory can be deleted. (select this
option from the PC menu.)

6.

When the transfer is complete, switch the analyzer off.

Software Installation

The internal battery can sustain 8 hours of continuous oper-
ation (approximately 100 tests) after charging for 24 hours. If
the battery becomes completely discharged, charge it for at
least 15 minutes before starting a test. The analyzer must
remain connected to the power adapter until the battery has
had time to recharge.

To recharge the battery, connect the lead from the power
adapter to the DC power input socket on the analyzer. Make
sure the battery charging indicator on the analyzer lights up.

Low Battery Level Indicator

When the low battery level indicator flashes on
the LCD display, recharge the battery as
soon as possible.

Important:

Before recharging the battery, always press the

red button to switch off the analyzer.

Conserving Battery Power

To help extend battery run time, use the power adapter
whenever possible. You can also save battery power by
switching the analyzer off between samples, and by
turning off the Auto Print mode. The LCD screen backlight
dims automatically when the external power supply is not
connected.

Battery Charging

Press the log button [6] to retrieve results s t o r e d
in the analyzer’s memory; the Log Screen will be
displayed.

1) Transfer Log

2) Clear Log

3) Clear Last

4) Recall #0

5) Print

Press a Key to Choose or 0 to Exit

Log Screen

Memor y Recall

Содержание Vickers Target-Pro 2

Страница 1: ...Vickers by Danfoss Target Pro 2 Analyzer and Sampler Operation Manual...

Страница 2: ...and Sampler Operation Manual V PP MC 0002 E July 2023 Target Pro 2 Laser Particle Analyzer User Guide September 2004 Contents Operating Precautions 2 Introduction 3 ISO Cleanliness Code 3 Target Pro 2...

Страница 3: ...d using mineral oil as the operating fluid The analyzer uses the light extinction principle in which two lasers are projected through the fluid and toward photo diodes Particles that interrupt the bea...

Страница 4: ...5 to program the analyzer The display will show the Operations Screen To change analyzer settings follow this sequence 1 Test Reference Press 1 enter your reference details up to 16 characters and pre...

Страница 5: ...previoustest and elimi nates cross contamination between samples 8 Press the flush valve push button to close the flush valve pushbutton light goes off Alternate method Proceed to step 9 pressing the...

Страница 6: ...essel vented to the atmosphere To conserve battery life connect the analyzer to the power adapter for continuous operation Continuous Sampling Basic Operation 1 Follow instructions 1 5 on page 4 to se...

Страница 7: ...ogging Press 4 once to log every test repeat press 4 again to deselect logging Note If you deselect log every test the analyzer will only store the test results when the Target Cleanliness Level is ac...

Страница 8: ...n appropriate solvent such as mineral spirits before taking a sample Do not use acetone Clean sample bottles and associated products following the instructions on page 13 2 Draw at least 80ml of fluid...

Страница 9: ...a minimum 1 Half fill the sampling bottle with the fluid to be analyzed 2 Re cap the bottle and shake it vigorously for 30 seconds 3 Discard the fluid following current COSHH legislation 4 Repeat step...

Страница 10: ...from the PC menu 6 When the transfer is complete switch the analyzer off Software Installation The internal battery can sustain 8 hours of continuous oper ation approximately 100 tests after charging...

Страница 11: ...press the flush valve and see if fluid passes into the waste disposal bottle Body text Note If you suspect high water aeration levels contact Danfoss for advice Warning If excessive system contaminat...

Страница 12: ...roscope counting examines the particles differently from APCs the code is given with two scale numbers only These are at 5 m and 15 m equivalent to the 6 m c and 14 m c of the APCs ISO 4406 Cleanlines...

Страница 13: ...ers These relate to the use of the correct viscosity mineral fluid An even cleaner level may be needed if the operation is severe such as high frequency fluctuations in loading high temperature or hig...

Страница 14: ...000 25 23 17 100 000 CorrelationbetweenParticleSizesObtainedusingACFTDandNIST CalibrationMethods Particle Size Obtained Using ACFTD SIZE ISO 4402 1991 ISOMTD NIST ISO 11171 size m m c 1 4 2 2 4 6 3 5...

Страница 15: ...required by ISO 9000 QS9000 and similar quality management sys tems A procedure for determining the performance of APCs means the user can set minimum acceptablelevels Benefits include improved calib...

Страница 16: ...Phone 45 7488 2222 Danfoss Power Solutions US Company 2800 East 13th Street Ames IA 50010 USA Phone 1 515 239 6000 Danfoss Power Solutions Trading Shanghai Co Ltd Building 22 No 1000 Jin Hai Rd Jin Q...

Отзывы: