Electrical connection
5.2 Remote transmitter only: Electrical connection of sensor cable
5.2
Remote transmitter only: Electrical connection of sensor cable
1. Wiring the transducer cables.
The system SONO 3500 CT transmitter is supplied with 4 separate transducer coaxial
cables for 2 paths.
The cables can be used on all transducers, but can be marked for a specific transducer
(cables are not paired with a specific transducer). The cables are manufactured with
crimp on the cable ends for the connection board.
Connect each transducer cable to the sensor. Tighten the transducer cable glands to
ensure they are firmly sealed! Recommended torque for each sensor gland is approx. 20
Nm.
Note
Shortened cables
If the cables are shortened, all cables must be shortened equally, and the factory set
cable length must be changed accordingly in the transmitter.
NOTICE
Verification sealing
For SONO 3500 CT systems the cable length setting is protected by a verification
sealing. The verification sealing can only be broken by the user with the acceptance of
the local authorities.
2. Snap out the connection plate and loosen the grounding wire.
28
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
Содержание SONO 3500 CT
Страница 1: ...Ultrasonic Flowmeter Operating Instructions SONO 3500 CT 07 2019 Edition ...
Страница 2: ......
Страница 8: ...Table of contents 6 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 14: ...Introduction 1 5 Further Information 12 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 38: ...Electrical connection 5 5 Sealing of SONO 3500 CT 36 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 50: ...Commissioning 6 4 Commissioning via PDM 48 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 54: ...Functions 7 4 Hardware key 52 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 64: ...Service and maintenance 8 7 Battery disposal 62 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...