unwind the tubing from the insertion device by
pulling it carefully upwards.
CAUTION:
Do not touch the Dana inset
TM
II
introducer needle when unwinding the tubing.
Be careful not to pull too hard when the tubing is
fully released. You could accidentally remove
the infusion set from the introducer needle.
8
Connect Dana inset
TM
II to a pump reservoir
and the pump. Prime the infusion set. Ensure
there are no air bubbles in the tubing. Consult
pump manufacturer’s instructions for use.
CAUTION:
When priming the infusion set, hold
Dana inset
TM
II with the needle pointing down to
ensure that medication does not come in
contact with the adhesive backing paper.
9
Pull up gently to remove the adhesive backing
paper.
10 Prepare the inserter:
Place fingers on lined
indentations. Press lined indentations on each
side. Pull spring until you hear a “click”.
11
Carefully remove the needle guard by gently
twisting and then pulling it. Check that the soft
cannula does not exceed the introducer
needle.
CAUTION:
Before positioning Dana inset
TM
II
over a clean insertion site, place the tubing in
the slot to ensure that the tubing is not caught
underneath the infusion set during insertion.
12 Insert Dana inset
TM
II:
Position Dana inset
TM
II
over the insertion site. Simultaneously press the
round indentations on either side of the insertion
device to insert Dana inset
TM
II.
13
Push gently on the center of the inserter to
secure the adhesive on the skin.
14
Remove the serter/introducer needle by
grasping the center and pulling it gently back.
Gently massage the adhesive tape onto the
skin.
Fill the empty space in the cannula with the
following amount of insulin:
U-100 insulin:
6 mm : 0.1 units (0.001 ml)
9 mm : 0.2 units (0.002 ml)
15
Push the lid back in place until you hear a
“click”.
A
Dana inset
TM
II allows you to temporarily
disconnect your pump without having to
change the infusion set. Hold the adhesive in
place by placing a finger just in front of the
cannula housing, and gently squeeze the sides
of the connector. Pull the connector needle
straight out from the cannula housing.
B
Insert the disconnect cover into the cannula
housing until you hear a “click”.
C
If necessary, prime the infusion set until
medication dispenses from the needle.
CAUTION:
When priming the infusion set, hold
Dana inset
TM
II with the needle pointing down.
D
Place a finger on the adhesive tape just in front
of the cannula housing.
E
Push the connector needle straight in until you
hear a “click”.
E N G L I S H