Rad-icon 1520 GigE User Manual
33
Teledyne DALSA
Snapshot Mode
In this mode the detector will initiate a sensor reset at the rising edge of the trigger input
and succeeding to this it will read-out the sensor at the falling edge.
The sensor reset is actually a normal read-out of the sensor but the data is discarded.
The trigger output indicates the Exposure Time when no reset or read-out of the sensor
occurs.
To allow sufficient time for the sensor reset, the delay between rising edge of the trigger
input and when the detector is exposed to x-ray should equal or higher to the reset time
of the sensor (D).
Row
#
Time
Reset
Integration
Time
Frame
(
N
)
Frame
(
N
+
1
)
Trigger Input
Trigger Output
Sensor Data Output
A
C
Pixel Array
1
2
Readout
D
B
F
Full
Resolution
Binning 2x2
ROI
A: Integration Time
Trigger input high
B: Read-out Time
40 ms
17.5 ms
varies
C: Exposure Time = Trigger Output
A - D
D: Reset Time
B
E: Minimal Trigger In high
> D
F: Minimal Trigger In low
> B