28
Rad-icon 1520 Detector
Teledyne DALSA
Detector Control
The Rad-icon Detector controls, as shown by CamExpert, groups the detector-specific
parameters. The features which are underlined link to the paragraph which describes
these features in more detail.
Display Name
Feature
Acc Description
ReadOutMode
RW Select the way the sensor is read-out
Full Resolution
FullResolution
Read the full array of the image sensor
Binning 2x2
Binning
Bins in sensor vertically and digital in FPGA for
horizontally direction
Region of Interest
ROI
Select a region in the full image array
Trigger Mode
RW
Controls by what source initiate and how the
sensor read-out
Free Running
FreeRunning
Detector uses internal timing
Triggered
ExtTrigger
Read-out of sensor occurs on falling-edge of
trigger input
Snapshot
Snapshot
Sensor reset occurs on rising edge and sensor
read-out occurs on falling-edge of trigger input
Extended Exposure (us) ExtendedExposure RW
Time in microseconds between two successive
read-outs in FreeRunning Trigger Mode
Offset
Offset
RW
Sensor dark ADC offset level; based on full
scale of sensor
0%
Offset0pct
0
1.95%
Offset1pct95
1,95% of 16384
3.9%
Offset3pct9
3.9% of 16384
6.25%
Offset6pct25
6.25% of 16384
Full Well Selection
FullWell
RW
Selects the full well of the pixels in the sensor
array
High Full Well
HighFullWell
Large full well capacitor
Low Full Well
LowFullWell
Small full well capacitor (default)
ROI Start V
ROIStartV
RW
Vertical start position for ROI mode; min 0, max
1516, multiple of 2
ROI Stop V
ROIStopV
RW
Vertical stop position for ROI mode; min 31,
max 1547, multiple of 2
ROI Start H
ROIStartH
RW
Horizontal start position for ROI mode; min 0,
max 1516, multiple of 4
ROI Stop H
ROIStopH
RW
Horizontal stop position for ROI mode; min 31,
max 1547, multiple of 4