Rad-icon 1520 Detector
Teledyne DALSA
Triggered (External) Mode
In this mode the detector will take a single image at the moment a trigger signal is
applied to the detector. This mode is equal to the Snapshot mode but no reset is needed
before integration can start.
When using pulsed X-rays, the X-ray exposure including X-ray rise- and fall-times have
to be within the “Exposure Time” window (C).
The trigger output indicates the Exposure Time window when no read-out of the sensor
occurs.
Row
#
Time
ReadOut
Integration
Time
Frame
(
N
)
Frame
(
N
+
1
)
Trigger Input
Trigger Output
Sensor Data Output
A
C
D
Pixel Array
E
1
2
B
Full
Resolution
Binning 2x2
ROI
A: Integration Time
Trigger input period
B: Read-out Time
40 ms
17.5 ms
varies
C: Exposure Time
> 21 ms
> 12.5 ms
A - B
D: Minimal Trigger In high
> 1 ms
E: Minimal Trigger In low
> 61ms - E
> 30ms - E
B + C - E