30
•
Camera Performance and Features
Measuring Line Rate (Trigger)
See Camera Control Category in Appendix A for GenICam features associated with this section and
how to use them.
Related Feature:
The Measured Line Rate command is used to read the line (trigger) rate being applied, externally
or internally, to the camera.
Maximum Line Rate
The maximum line rate that the camera can achieve is determined by the number of CLHS lanes
used and by the number of cables installed, as shown in the following table:
Maximum Line Rate (1 sensor line output, 8 bit)
Camera Model
One Fiber Optic Cable
Two Fiber Optic Cables
One AOC cable
ML-FM-08K30H-00-R
140 kHz
280 kHz
NA
ML-FM-16K07A-00-R
70 kHz
70 kHz
NA
ML-FM-16K15A-00-R
71 kHz
143 kHz
NA
ML-HM-08K30H-00-R
NA
NA
300 kHz
ML-HM-16K30H-00-R
NA
NA
300 kHz
Minimum Line Rate
The minimum line rate for all camera models is 0 kHz. Cameras go to full stop without image
anomaly.
Scan Direction
See the section Camera Control Category in Appendix A for GenICam features associated with this
section and how to use them
Related Feature:
A multiline camera requires the user to tell the camera the direction of travel of the object being
imaged.
The scan direction is set using the sensorScanDirectionSource command. The options are: Internal,
Line 2 (pin 6 on the GPIO connector), and the rotary encoder feature (using pin 5 and pin 6 of the
GPIO connector).
When set to internal, use the sensorScanDirection feature to set the direction.