Falcon™ 4
-CLHS Series
Technical Specifications
•
63
Sensor Alignment Specification
The following figure specifies sensor alignment for Falcon4 where all specifications define the
absolute maximum tolerance allowed for production cameras. Dimensions
“
x, y, z
”
, are in microns
and referenced to the Falcon4 mechanical body or the optical focal plane (for the z-axis
dimension). Theta specifies the sensor rotation relative to the sensor
’
s center and Falcon4
mechanical.
X variance
+/- 100 microns
(+/-
)
Y
variance
(+/-)
X
variance
Sensor Alignment Reference
Z
variance not shown
(+/-)
theta
variance
Y variance
+/- 100 microns
Z variance
+/- 300 microns
Theta variance
+/- 1 degree