Fa
lcon™ 4
-CLHS Series
Operational Reference
•
37
Step 6.
In the Calibration FPN feature, click the
Press…
command to perform the calibration.
Step 7.
If required, verify the Optical Black Reference Calibration Offset setting; set the FPN
Grab an image and use a histogram to verify that pixel values are above zero; if necessary, adjust
the Optical Black Reference Calibration Offset. Higher offset values may be required when the
camera is operating at very high temperature and frame rates. Set the FPN Correction Mode to
Calibration and use the Calibration FPN command to recalculate the calibration and adjust the
offset until the result is satisfactory.
Step 8.
If the calibration is satisfactory, in the Save FPN Calibration feature click the
Press…
command to save the calibration to the selected user set.
For each gain setting required perform and save the calibration; coefficients for each gain setting
are saved in the user set (gain settings for which no calibration is done use default coefficients).
Step 9.
To load this user set when resetting or powering on the camera, use the Power-up
Configuration dialog. Click
Setting…
Power-up Configuration feature, in Camera Information
category.
In the
Power-up Configuration
dialog box select the required user set both
Camera Power-up
configuration
and
Load
/
Save configuration
and click Save.
This ensures that the camera loads the saved parameters the next time the camera is turned on.
Содержание Falcon 4-CLHS Series
Страница 13: ...Falcon 4 CLHS Series Falcon4 CLHS Specifications 11 Quantum Efficiency Curves Spectral Responsivity ...
Страница 14: ...12 Falcon4 CLHS Specifications Falcon 4 CLHS Series Effective Quantum Efficiency ...
Страница 87: ...Falcon 4 CLHS Series Additional Reference Information 85 Cable Assembly G5 AIOC BLUNT2M ...
Страница 88: ...86 Additional Reference Information Falcon 4 CLHS Series ...
Страница 89: ...Falcon 4 CLHS Series Additional Reference Information 87 Generic Power Supply with no I O ...