Performing an FPN Setup via Sapera CamExpert
The Sapera LT CamExpert tool provides an easy GUI based method for a user to perform a FPN
Calibration. For FPN calibration the camera must acquire a suitable dark image.
FPN Correction
In general, factory FPN correction is sufficient for most applications. However, if a new FPN
correction is required a user set FPN correction can be applied.
Important:
Before calibration, the Falcon4 should be powered on long enough to
achieve its nominal temperature (a minimum of 30 minutes). A low ambient
temperature may increase the time required for the camera to reach a stable
internal temperature.
Important:
During calibration, no other Falcon4 features should be accessed or
modified.
Calibration via CamExpert or via a User Application:
Exposure and frame
rates used during a Flat Field Calibration should be similar to the exposure settings
used in the camera application.
To perform FPN Correction
:
Step 1.
Cover the lens (place the sensor in dark).
Step 2.
Set the FPN Correction Mode to Off and check the histogram.
Using CamExpert, click
Grab
and then
Statistics
.
In the Statistics dialog, use the
Selected view
drop-down list to select Histogram.
Содержание Falcon 4-CLHS Series
Страница 13: ...Falcon 4 CLHS Series Falcon4 CLHS Specifications 11 Quantum Efficiency Curves Spectral Responsivity ...
Страница 14: ...12 Falcon4 CLHS Specifications Falcon 4 CLHS Series Effective Quantum Efficiency ...
Страница 87: ...Falcon 4 CLHS Series Additional Reference Information 85 Cable Assembly G5 AIOC BLUNT2M ...
Страница 88: ...86 Additional Reference Information Falcon 4 CLHS Series ...
Страница 89: ...Falcon 4 CLHS Series Additional Reference Information 87 Generic Power Supply with no I O ...