8. Sterilization
1) For cleaning and sterilizing, the unit must be disconnected from the mains. Accessories
have to be always keep sterile when you use in operating room, especially metal parts which is
contacting with the patient's body.
2) Be sure to clean and sterile the accessories before keeping them.
3) Before sterilization, inspect the accessories and their cord for breaks or cracks in the insulation,
corrosion, or loose conditions. Do not use the accessories if any of these conditions exist.
4) Sterilization
1. Do not overheat or overpressure or overtime, it can damage the accessories.
2. Do not clean or sterilize disposable accessories. The disposable accessories should be used only
one time.
3. Accessory life will be reduced:
- High mineral content in the autoclave supply water.
- Tight bunching or wrapping of the cord during sterilization.
- Contact between the cord and material instruments during autoclave sterilization.
4. Sterilize the accessories made of plastic, PVC or rubber with EO gas as possible. If they are in
sterilize with autoclave, they may be transformed or the life span may be reduced due to high
temperature.
5. The main body of the unit can not be sterilized.
6.Recommendation on the use of non-flammable agents for cleaning and disinfections wherever
possible
Accessory
Sterilization
Steam
(Autoclave)
EtO
(EO Gas)
1. Twin Button Handle with Cable
2. Mono-polar Handle with Cable
3. Forceps Cable
4. Bi-polar Forceps
5. Electrodes
6. Reusable Return Plate Cable
ㆍ
Gravity
displacement
ㆍ
121
℃
(250
℉
) for
20minutes
ㆍ
Concentration of EtO
: 600mg/liter
ㆍ
Temperature
: 54
℃
(130
℉
)
ㆍ
Exposure time
: 1 hour 45 minutes
ㆍ
Aeration time
: 12 hours at 54
℃
(130
℉
)
7. Reusable Metal Plate with Cable
Non-available
8. Reusable Rubber Plate with Cable
Non-available
Caution
Содержание Meditom DT-400S
Страница 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Страница 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Страница 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Страница 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Страница 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Страница 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Страница 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Страница 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Страница 47: ...9 Load Regulation Vpeak Graphs47 9 5 Forced Soft Coagulation Mode 9 5 1Load Regulation DT 400S DT 300S ...
Страница 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...