4
. Effect using the unit27
- Can coagulate the tissue wetted by blood or any other liquid.
2) BipolarSoft Coagulation
- 482KHz alternating current maximum modulated by 20KHz.
- Minimize tissue adhesion by
‘Time restricted output’.
- Especially good to small tissue coagulation.
- Select this mode by 'Bi-polar Coagulation Mode Select Switch[16]'
- 'Bi-polar Coagulation Mode Indicator[17]' illuminates(not blink) in this mode.
TIP How to apply bi-polar mode
- This mode is useful for operations on sensitive tissues such as nerve or blood vessel. Those who
have not much skill should be fully aware of the following checkpoints.
- It often happens to deal the tissues again since they are subject to stick to the tips of Forceps. You
can minimize the problem above mentioned by using the following procedures.
- The problem occurs because the current going through the Forceps heats its tips and this makes the
tissues adhering to the tips dehydrated too much. To solve this problem, it is essential to reduce the
heat at the tips. Follow the procedures below.
- Don't supply HF current (output) to the tissues for a long time.
- Use the minimum output just so as to make cutting and coagulation. If you supply higher output than
is necessary, the tips of Forceps will be too much heated, coming out the gilt of tips. If it is not
evitable to use high output, be mindful to cool the tips sufficiently.
- Use the Forceps with the wide tips for coagulation.
- Use the Forceps with the narrow tips for cutting.
- Replace Forceps with new one from other manufacturer as the quality and processing method vary
according to the manufacturers. Low quality Forceps will often cause the case as stated above.
- Check the silt status at the tips of Forceps since the peeled silt from tips or coarse Forceps makes the
condition worse. Especially, be careful to wash the Forceps not to damage the tips. If some tissues
are adhered to the tips, remove them with a piece of soft cloth after soaking the Forceps in the water.
- Don't remove liquid from the tissues on purpose because the tissues without enough liquid are subject
to adhere to Forceps. Supply some water such as saline solution to dry tissue to avoid this problem.
TIP What is
‘Time restricted output’?
- This method means how to supply output by attenuating it based on the time. See the figure
below.
- You can use this method when dealing tiny tissues as it strongly avoids carbonization and
adherence of tissues.
- The large tissue may not be fully coagulated by supplying output once. Accordingly, press the
foot switch to supply the output once again or change the mode to Standard Coagulation one.
Содержание Meditom DT-400S
Страница 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Страница 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Страница 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Страница 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Страница 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Страница 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Страница 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Страница 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Страница 47: ...9 Load Regulation Vpeak Graphs47 9 5 Forced Soft Coagulation Mode 9 5 1Load Regulation DT 400S DT 300S ...
Страница 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...