DAIHAN Scientific, Steam Sterilizers Digital Fuzzy-controlled Autoclaves .
23/53
* After Sterilization
-
If the sterilization operation is completed, all steam is safely emitted, and if the Open the
Door Message is appeared on the LCD Display along with Buzzer sound, the user can safely
take out the Sample. (If less than 80
℃
Message appears.)
-
Door Open Message Display will continue until the Open Door.
(when the Pressure "0", temperature "80
℃
" "Door Open" message will appear. Please be
careful it may be hot in the internal temperature.)
-
After checking that the pressure of the mounted pressure gauge is zero (0), press the Open
button and please open the door by pulling the Lever.
-
To check the perfect pressure ejection of the pressure vessel before opening the door of the
pressure vessel, the user open the Manual purge valve and can finally check.
-
If the user wants to remove the remaining distilled water in the pressure vessel, it can be
removed by opening the Water Drain Valve. Please check the device which will collect the
drained distilled water, before opening the valve.
* Error Message
* Emergency Lid unlock method
autoclaves has electronic lid lock system for customer’s safety. So, if customer face power failure
or something wrong in power supplier of units, customer couldn’t open the lid by push the
“Open” button. In this case, for opening the lid, please do below procedure.
ERROR Message
Cause
Remedy
Error 1
Informs the time to be displayed the
temperature of abnormal value due to
the abnormality of Temperature Sensor.
.
Please check the setting
temperature.
Error 2
The mounted inner water level
sensor informs to user when the
distilled water is not filled up to
Heater.
Please check the amount of
internal distilled water.
Error 3
Inform when can not operate owing
to Opened Door.
Please check the amount of
internal distilled water.
Temp SV : 120
℃
Temp PV : 79
℃
Temp SV : 120
℃
Open the Door !!!
< Fig 3.3.11. Door Open Message>