                                                                  8/53 

2. Installation         

                                                                                                             

 

      2.1 Product Installation   

* Caution of Installation 

1.

 

Please be careful when transporting the product because it is heavy. 

2.

 

Please do not lay the product on the floor or subject it to shock. It can cause damage inside 

the product. 

3.

 

Please install the product in a hard, flat, and stable place. 

4.

 

Please do not install the product in a place where it is exposed to direct sunlight, or in the 

dangerous places. 

5.

 

To minimize the risk of short circuits, do not install the product (in the) places where it will 

be subjected to the inflow of moisture or organic solvents on the regulator part or inside of 

the body. 

6.

 

This product was manufactured for normal operation at a rated voltage, so please check the 

voltage status before installing the product. 

7.

 

When installing and using the product, please certainly use the power cord provided with 

the products. 

* Installation Process 

1.

 

After receiving the unit and putting it on the specified place of use remove the wrapping. 

2.

 

After positioning the unit, lock the moving wheel in order to make position fixed and the 
unit stable. 

3.

 

Check the unit whether there are eventual transport damages or not. In case there are 
some transport damages found please inform the forwarder and your contractor. If there is 
no damage, please go to next point 3. 

4.

 

Let the unit stand for a few minutes so that the all the internal parts and components of the 
unit can be stabilized. 

5.

 

Open  the  door  at  the  left  side  of  the  unit.  You  can  find  a  room  for  pipe  assembly  and 
electric  distribution  plane.  Please  check  carefully  if  there  is  anything  wrong  including 
disconnection of the electric wire, breakage of pipe or fitting element. In case there is any 
kind of damage found, please inform the forwarder and your contractor. If you cannot find 
anything wrong, please go to next point 5. 

6.

 

For normal use of the unit, confirm that the valve is open, and water drain valve is closed. 

7.

 

Connect the main power cord to the socket.   

8.

 

(NOTE)

 If you bye MaXterile

TM

 80R or MaXterile

TM

 100R, power consumption of the heater 

in  the  vessel  is  too  high  to  use  normal  wall  socket.  (Typical  wall  socket  in  EU  and  Asian 
countries permits around 3kW maximum per socket) So, a big 3-pole plug (for single-phase 
230V  AC  with  ground)  comes  with  power  cord,  and  a  socket  with  wiring  connector  is 
supplied together as seen in the picture on the page 10. You have to wire the power supply 
line  of  enough  electric  power  (4kW  for MaXterile

TM

  80R  and  5kW  for MaXterile

TM

  100R  at 

least) to the supplied connector before plugging. 

9.

 

If such a big power supply line is not available in the room where the unit is installed, you 
have  to  contact  the  appropriate  person  who  is  in  charge  of  electric  facilities  in  your 
company or organization. 

Содержание MaXterileTM 47

Страница 1: ...dation IQ OQ Service Patented No 10 0554098 Jog Shuttle Control System User Manual Version 2 2 1 written for MaXterileTM 47 60 80 100 MaXterileTM 60 Steam Sterilizers Digital Fuzzy controlled Autoclav...

Страница 2: ...rol with monitoring of final assessment Category II Autoclave 1 Item Vessel of Autoclave 2 Model names MaXterile 47 MaXterile 60 MaXterile 80 3 Design Pressure 2 5bar 4 Operation Pressure 1 2 2 bar 5...

Страница 3: ...product Autoclave MaXterile 100 to which this declaration relates is in conformity with the following standard or other normative document EN 61326 EN 61000 3 2 EN 61000 3 3 EN 61010 1 EN 61010 2 foll...

Страница 4: ...ore Use 5 1 2 Product Introduction 6 1 3 Product Configuration 7 2 Installation 2 1 Product Installation 8 2 2 Product Specifications 10 3 Product Usage 3 1 Product Usage 12 3 2 Names of Each Part 15...

Страница 5: ...well acquainted with the meaning Warning Failure to follow the instructions may result in injury Caution Failure to follow the instructions may result in product failure Proximity Caution Keep clear...

Страница 6: ...ess steel material processed by a press is used in order to maintain the aesthetics of Chamber and the high level of hermeticity and the outside overheating of the equipment is protected against in th...

Страница 7: ...igital Fuzzy controlled Autoclaves 1 Unit Wire Basket Set 2 ea Heater Cover 1 ea Drain Bucket 1 ea Including a copy of the Operation Manual Model MaXterileTM 47 Steam Sterilizers Digital Fuzzy control...

Страница 8: ...e inform the forwarder and your contractor If there is no damage please go to next point 3 4 Let the unit stand for a few minutes so that the all the internal parts and components of the unit can be s...

Страница 9: ...nsurance so that its customers can trust our products believe and use its products by preparing against the accident which can be occurred during using the products by self production Contents Insuran...

Страница 10: ...Function Controller Display Fuzzy Control System Fully Automatic Door Closing System Lever type Sliding Door Safety Electronic Door Closing System Safety Circuit Over Temp Pressure Current protection...

Страница 11: ...DAIHAN Sci com or ALL forLAB com Shopping Mall DAIHAN Scientific Steam Sterilizers Digital Fuzzy controlled Autoclaves 11 53 Functional Block Diagram...

Страница 12: ...esigned for the following purposes For sterilizing glassware rubber items pipette tips instruments filter bodies as well as consumables waste in bags and to destroy infectious or septic materials For...

Страница 13: ...stone a metal or paper goes insides it may cause a fire an electric shock or a break down If do pull out the mains plug from a socket and contact a manufacturer or a supplier Do not disassemble or re...

Страница 14: ...s or instruments made from steel or iron Metering equipment unless their suitability for stream sterilization is specifically mentioned Do not open sterilizing chamber until full sterilization operati...

Страница 15: ...ge Breaker Circuit Breaker Power Switch Power ON OFF Lever Door Safety Lock and easy open Lid Clamp Door Locking Control Panel LCD Display Pressure Gauge manometer Moving Wheel Transport Device Table...

Страница 16: ...Names Description Door Fastener Door locking mechanism Safety Knob Prevent overheat Manual purge valve pressure pass Control Device Water Drain Valve Drain Control Device Condenser emission steam con...

Страница 17: ...of TIMER HEAT ERROR MAIN Button Convert into mode of Timer Heating Printer Locking SUB Button Convert into mode of delay timer offset function Setting Operating Information OPEN Door open Jog Dial kno...

Страница 18: ...turning the Jog Dial knob to clockwise or counter clockwise the setting values can be adjusted and if briefly pressing once the PUSH button the Start or the Stop operation is performed The Turning of...

Страница 19: ...y setting the reservation time value with the Jog Dial knob the value is fixed on the Display and the reservation timer is activated After setting the reservation timer value you can set up the operat...

Страница 20: ...blinking in long interval The range of the setting temperature is adjustable in the unit of 1 and can be set up to 132 You can change or save the setting by using the operation method of the Jog Dial...

Страница 21: ...itched to the Set Temperature Mode MAIN Safety Locking If pressing the MAIN button in the Set Temperature mode it is switched to the Safety Locking Mode In the Locking Mode the Turning by the Jog Dial...

Страница 22: ...he steam is rapidly and safely emitted through the emission steam condensation mechanism In LIQUID cycle mode valve is not completely opened at a time Firstly it waits for 5 minutes with valve closed...

Страница 23: ...eck If the user wants to remove the remaining distilled water in the pressure vessel it can be removed by opening the Water Drain Valve Please check the device which will collect the drained distilled...

Страница 24: ...nagement 4 1 Accessories Ordering Information Image Ordering Number Information DH WAC20047 Perforated Basket 270 x h250mm WAC20047 for 47Lit MaXterileTM 47 MaXterileTM 47R DH WAC20060 Perforated Bask...

Страница 25: ...ay cause an electric shock or fire To clean the unit a neutral detergent and soft cloth is recommended Do not use coarse cloth strong chemicals or organic solutions Do not use water Benzene Thinner or...

Страница 26: ...e of the device If the temperature does not rise at room temperature Check the temperature currently set Make sure that the power switch is in ON position Make sure that the Push button of the Jog Dia...

Страница 27: ...t This instrument is warranted for 2 years from purchasing date A breakdown caused by customer s improper use or natural disaster is excepted from this warranty This certification is only for internat...

Отзывы: