DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Forced Convection type Smart Oven 24/80
<Fig.3.3.4> set graph display cycle
<Fig.3.3.5> reserve opeation setting
B) Program Mode: It restarts from the first segment of the previous pattern it operated on.
-
Setting a Continue mode: It operates depending on (4) Operation Mode.
A) Fixed Mode: When reauthorizing the power, it automatically operates at a
temperature set that it previously operated on and the operation time on main
screen continues its counting.
B) Program Mode: It restarts the count from the remaining time of first segment that
stopped and it restarts.
(2)
Chart Display Cycle
-
You can set the temperature cycle shown on Chart. User can select from 30 seconds, 1
minute and 2 minutes.
(3)
Reserve Mode
-
Set date and time you want to operate the device and it will start at the reserved time.
Minimum time required to use this mode is 5 minutes.
-
It automatically turns off after being applied one time.
(2)
(3)