Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The DTS-3000 uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The DTS-3000 is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply net work that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Information on EMC
9-1. Guidance and manufacturer’s declaration
- electromagnetic emissions
9-2. Guidance and manufacturer’s declaration
– electromagnetic immunity
The DTS-3000 is intended for use in the electromagnetic environment specified below.
The customer or the user of DTS-3000 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagneticenvironment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for power
supply lines
Mains power quality should be that
of a typical commercial or hospital
environment.
The DTS-3000 is intended for use in the electromagnetic environment specified below.
The customer or the user of DTS-3000 should assure that it is used in such an environment.
9
39
UserManual_DTS-3000_170223_인증용.indd 39
2017-02-23 오후 2:06:10