10
2CH
Deflate
Button
Deflate the cuff of 2CH
For deflating a cuff, press the Deflate button long (3seconds).
LCD displays during deflating.
11
2CH Cuff
Change
Button
Change the hemostasis area while using
double cuff of 2CH
When operation continues for a long time, change the
hemostasis area while using double cuff to prevent damage of
skin, nerve or muscle.
Press the Cuff Change button long (3 seconds) while using the
Cuff A. Pressure is applied to the Cuff B and then the Cuff A is
deflated. In this case, LCD shows the change from the Cuff A
to the Cuff B at the bottom of the screen.
12
2CH LOP
Button
Set LOP of 2CH
LOP notifies the proper pressure for hemostasis by measuring
of a patient’s LOP.
For using LOP, press the LOP button long (3 seconds) after
putting a cuff on, and LOP measurement starts. Then, LOP
automatically estimates and notifies the proper pressure for a
patient. To start, press the Inflate button to activate. The icon
below is activated on the LCD during LOP
measurement. LOP icon color is changed as shown below.
13
Record
Button
Record the history of application
Setting values or errors are saved during the application.
14
Power
Button
Power Button
Turn power on or off with this button.