2
1CH
Inflation
Button
Inflation of the cuff of 1CH
Press the Inflate button long (3 seconds) to apply preset
pressure to a cuff.
When pressure is applied to a cuff, LCD displays .
When pressure reaches to the preset value, is
displayed.
3
1CH Time
Setting
Button
Time setting of 1CH
The previous value is saved and displayed. It can be adjusted
to 1 to 240 minutes depending on patient’s status or operation
types. Press the Time Setting button ( , ) to set the time.
User can change the set pressure during operation. (Unit :
5min (Long push), 1min (Short push)). The time displayed
during operation indicates the elapse time. Time can be
adjusted during operation using the button ( , ).
4
1CH
Deflate
Button
Deflate the cuff of 1CH
For deflating a cuff, press the Deflate button long (3seconds).
LCD displays during deflating.
5
1CH Cuff
Change
Button
Change the hemostasis area while using
double cuff of 1CH
When operation continues for a long time, change the
hemostasis area while using double cuff to prevent damage of
skin, nerve or muscle.
Press the Cuff Change button long (3 seconds) while using the
Cuff A. Pressure is applied to the Cuff B and then the Cuff A is
deflated. In this case, LCD shows the change from the Cuff A
to the Cuff B at the bottom of the screen.
6
1CH LOP
Button
Set LOP of 1CH
LOP notifies the proper pressure for hemostasis by measuring
of a patient’s LOP.
For using LOP, press the LOP button long (3 seconds) after
putting a cuff on, and LOP measurement starts.