Display
Test method
The test starts from CH1 after inflating for
10 seconds whether there pressure stays or
not. If pressure sensor does not recognize
pressure, then the machine decides there is
problem with Solenoid and shows solenoid
error.
The machine does not perform following test.
Sensor check / Leak check test.
Display
Test method
- It tests if pressure can reach to the cuff for
30 seconds in case the pressure sensor is
recognized pressure
- If the pressure reach to the specific
pressure figures that selected by user within
30 seconds then it stops motor and stay. It
stays 30 seconds and it decides if there is
any leak.
It stop testing and test next chamber in
case there is some problem.
Display
Test method
- After completing test until 4CH then it
automatically stop all.