9. Product Inspection
Symptom
Cause
Action
When Display
trembles.
•
Load cell is damaged.
•
Load cell insulation
resistance.
•
Indirect occurrence
•
Load cell input, output.
•
Check resistance
•
Check load cell’s insulation resistance.
When weight goes
up at a regular ratio
or zero returns are
not made.
•
Loadcell faulty
•
Check load cell’s insulation resistance.
•
Loadcell connection is
insufficient.
•
Check the wiring between load cell and the
main device.
•
Check the load cell’s calbe’s disconnection.
When weight
changes into (-).
•
Loadcell wiring is
reversed.
•
Check load cell’s ouput
cable connection.
Displayed as Lc_Er
•
Load cell is damaged.
•
Load cell connection is
bad.
•
Check the load cell’s condition and calbe
connection.
Displayed as c_Err
•
Actual calibration set
below zero (-)
•
Performing actual load calibration at plus(+)
If the number on
the Display does
not match
•
Out of maximum display
range(-19999 ~ 99999)
•
Re-check the 8-4 Digital Calibration setting
and the F-51 Force unit setting.
•
Restore to the settings saved by RESTORE
execution
Pressing
key
in FUNC state does
not move to
calibration mode.
•
Locked State at 8-7
Lock Set-up.
•
State in which units of
conversion at F-51 Force
unit is established.
•
Unlock from Locking Settings at 8-7 Lock
Set-up.
•
Change to kgf unit (set value :0) at F-51.
Force unit
The number of
measurements is
incorrect
•
Maximum Display
Range Out of 99999
•
Check calibration settings and F-51
settings. (page 30~36), (page 29)
•
Restore to the settings saved by RESTORE
execution (page 37)