29
F-41. Baud rate (Communication Speed Setup)
(Standard setup value : 9.60)
Display data
Setting
2.40
2400 bps
4.80
4800 bps
9.60
9600 bps
19.20
19200 bps
38.40
38400 bps
57.60
57600 bps
PRINT DATA OUT (PT-100)
F-42. Protocol (Communication Protocol set up)
(Standard setup value : 0)
Display data
Setting
0
Standard protocol
1
Modbus RTU Protocol
F-50. Display reverse mode (
Direction setting, mark reverse
)
(Standard setup value : 0)
Display data
Setting
0
Positive mark
1
Negative mark
2
Display as absolute value
F-51. Force unit (Conversion unit set up)
(Standard setup value : 0)
Display data
Setting
0
kgf
kg/
㎠
1
N
kgf × 9.8
-
-
2
ℓ
b
kgf × 2.2
-
-
3
-
-
Bar
kg/
㎠
× 0.98
4
-
-
MPa
kg/
㎠
× 0.098
Note1) When setting the conversion unit (set value: 1 to 4), calibration modes of 8-3
and 8-4 are not available.
After calibrating in kgf (set value:0), set the force unit as desired.
Note2) If F-51 is set to MPa (set value:4), the F-00 Decimal point setting cannot be
changed.