12
Distance protection against parasite radiation
The operator must use the safety device of the equipment (distance activation cable), remaining at
a minimum distance of 3 meters from the X-radiation beam during the charge application. The user
must wear the protection clothes (lead apron and lead collar) they are not provided with the product.
Lead apron and collar,
(They are not provided with
the product).
How to regulate arm tension
If the arm assembly/cylinder head loses its balance condition, a technician authorized by Alliage
shall be called in order to make the adjustment.
Radiation quality
Ingress of quality equivalence
Parametes for charge exposure 0,32s
FILTERS Tube crystal
1,26 mm Al
Insulating oil, column 17mm
0,64 mm Al
Window
0,32 mm Al
Additional filter (aluminum)
1,00 mm Al
TOTAL FILTRATION
3,22 mm Al
Tension
70 kVp ± 10%
Current
7 mA ± 1,5 mA
Power
0,49 kW
Product current time
2,24 mAs
Highest energy applied in intermittent
mode during 1 h
170 KJ
TECHNICAL SPECIFICATIONS
General features
Содержание Spectro 70X Series
Страница 1: ......
Страница 15: ...15 Dimensions mm Model Coluna M vel TECHNICAL SPECIFICATIONS...
Страница 16: ...16 Dimensions mm Model Parede TECHNICAL SPECIFICATIONS...
Страница 17: ...17 Dimensions mm Model Pantogr fico Coluna M vel TECHNICAL SPECIFICATIONS...
Страница 18: ...18 Dimensions mm Model Pantogr fico Piso TECHNICAL SPECIFICATIONS...
Страница 19: ...19 Dimensions mm Model Pantogr fico Parede TECHNICAL SPECIFICATIONS...
Страница 22: ...22 Content of accessible and non accessible demarcations 01 02 TECHNICAL SPECIFICATIONS...
Страница 23: ...23 03 04 05 TECHNICAL SPECIFICATIONS Content of accessible and non accessible demarcations...
Страница 24: ...24 06 07 08 10 09 TECHNICAL SPECIFICATIONS Content of accessible and non accessible demarcations...
Страница 42: ...42...
Страница 43: ...43...
Страница 44: ......