M5-CYG2P-M-01_Iss6.doc
Cygnus 2+ Operating Manual
57
Cygnus M5-C2P
Technical Specifications
Type of Display
OLED Monochrome
Display Size
128 x 32 Pixels. 25.58 mm (W) x 6.38 mm (H)
Display Information
Digital Thickness Value. Settings. Battery Level.
Transmitter
Shape of Pulse
Square
Pulse Energy : Voltage (peak-
to-peak)
70 V p-p
Pulse Energy : Rise Time
3 ns (max)
Pulse Energy : Pulse Duration S2C : 220 ns
S2D : 220 ns
S3C : 100 ns
S5A / S5C : 67ns
T2C : 220 ns
T5B : 100 ns
T7A : 67 ns
HT5 : 100 ns
Receiver
Gain Control
Automatic Gain Control depending on probe and measurement
mode.
Frequency Range
1.0 MHz to 10.0 MHz (-6dB)
Other Information
Data Output and Storage
Non.
Data Connector
USB Mini B Connector under battery cover.
Can be used to update the gauge firmware and load gauge
settings.
Calibration setting storage
Calibration Data stored to Internal Flash Memory
Calibration Mechanisms
Not required for Multiple Echo mode.
Automatic V-path correction for twin element probes.
Option of two point calibration for twin element probes.
Display & Recall Facilities
N/A
Display Response Time
125 ms / 500 ms
Printer Output
N/A
Environmental Rating
IPX67 (Water immersion 1 metre depth for 30 minutes)
MIL STD 810G Method 501.6 (High Temp +55°C)
MIL STD 810G Method 502.6 (Low temp -20°C)
MIL STD 810G Method 507.6 (Humidity 95%)
MIL STD 810G Method 512.6 (Immersion 1m, 30min)
Shock & Impact
MIL STD 810G Method 514.7 (Vibration)
MIL STD 810G Method 516.7 (Shock 20g)
MIL STD 810G Method 516.7 (Transit Drop 1.22m)
Compliance
RoHS Compliant.
Designed for BS EN 15317:2000.
CE Marked including EMC.