Cygnus 2+ Operating Manual
M5-CYG2P-M-01_Iss6.doc
34
...use the Up and Down Select
keys to adjust the thickness to
the required value.
8.
When the correct measurement
is displayed press the menu key
9.
Next measure the Thin sample –
Minimum thickness.
While holding the probe firmly
on the thin sample, and while a
steady thickness measurement
is displayed..
...use the Up and Down Select
keys to adjust the thickness to
the required value.
10.
When done press the Menu key
to save the calibration.
Or press the Power / Cancel to
exit without saving the
calibration.
Probe Zero (twin element probes)
Twin element probes must be zeroed to compensate for any wear
or operating temperature changes. The gauge will always perform