Maintenance
&
Care
9.2.4
Tip Magazine
Operating
the
CyBi
®
-Well
in
manual
or
automatic
mode,
you
should
carefully
watch
for
and
remove
residual
air
bubbles
from
tips.
Otherwise,
the
liquid
in
affected
tips
will
reach
an
excessively
high
fill
level.
There
is
danger
of
pistons
coming
into
contact
with
liquid
and
suffering
contamination
as
a
result.
This
may
happen
if
liquid
is
drawn
from
vessels
with
insufficient
filling
level
or
if
tips
are
not
submerged
deep
enough.
Make
sure
that
pistons
cannot
come
into
contact
with
liquid
during
motion
into
zero-
position.
Repeated
faulty
operation
may
lead
to
excessively
high
fill
levels
and
a
situa-
tion
of
the
kind
as
described
above.
CAUTION
Direct
contact
of
the
tip
magazine
and
the
pistons
may
damage
the
pistons.
Never
install
a
tip
magazine
without
tips
into
the
pipettor!
The
insertion
of
a
tip
maga-
zine
with
no
tips
installed
will
void
any
warranty
claims!
NOTE
Before
shutting
the
system
for
some
hours
or
some
days,
the
tips
should
be
unfixed.
The
tip
magazine
must
remain
installed
in
any
such
case.