c) After the inflation process lay down the pump
ball and remain in a quiet and relaxed sitting position. The
measurement now proceeds on its own.
If no measurement takes place and all that appears in the
display is a flashing, upwards pointing arrow, pump the ball
once again and inflate to a higher pressure «Inflate further».
When the instrument detects a pulse, a heart symbol in the
display starts to flash and a beep is heard for every heart
beat.
d) When the measurement has been concluded, a long beep
tone sounds. The measured systolic and diastolic blood
pressure values as well as the pulse are now displayed.
The cuff pressure can now be quickly released by pressing
the release valve at the pump ball.
5.5. Inflating further
If the cuff is not inflated sufficiently, the measure-
ment is stopped after a few seconds and a flashing
arrow pointing upwards appears in the display.
It is then necessary to inflate at least 20 mmHg
higher than the previous value.
The instruction to inflate further can appear several
times if the pressure is still not sufficient.
Attention:
If the cuff is inflated too far (over 300 mmHg), «HI»
appears in the display and a warning signal is
sounded. The cuff pressure must then be immedi-
ately released using the quick-release valve at the
pump ball!
As soon as the cuff is inflated sufficiently and the
measurement is proceeding, continue as described
under points c + d.
5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure mea-
surement for any reason, press the quick release
valve at the pump ball. The device immediately low-
ers the cuff-pressure automatically.
9
SYSTOLIC
mmHg
DIASTOLIC
mmHg
PULSE
/min
SYSTOLIC
mmHg
DIASTOLIC
mmHg
PULSE
/min
SYSTOLIC
mmHg
DIASTOLIC
mmHg
PULSE
/min
SYSTOLIC
mmHg
DIASTOLIC
mmHg
PULSE
/min
Содержание #271243
Страница 16: ......