17
1. The right ear reading may differ from the reading taken at the
left ear. Therefore, always take the temperature in the same ear.
2. The ear must be free from obstructions or excess earwax buildup
to take an accurate reading.
3. External factors may influence ear temperatures such as:
●
lying on one ear or the other
●
their ears covered
●
exposed to very hot or very cold temperatures
●
recently swimming or bathing
In these cases, remove the individual from the situation and wait
20 minutes prior to taking a temperature.
4. For persons wearing hearing aids or ear plugs, remove the device
and wait 20 minutes prior to taking a temperature.
5. When using the thermometer on infants under age 1, pull the
ear up making sure the sensor faces the eardrum.
(See Figure 5)
6. When using the thermometer on individuals over the age of 1,
pull the ear back making sure the sensor faces the eardrum.
(See Figure 6)
7. Do not touch the probe window. A cotton swab and rubbing
alcohol may be used to clean the sensor window.
8. If the thermometer is stored in a significantly different
environment than testing location, place it in the testing
location for approximately 30 minutes prior to use.
9. It is not intended for use in the oxygen rich environment and
presence of flammable anesthetic mixture with air, oxygen or
nitrous oxide.
15
16
Temperature Taking Hints
18
Temperature Taking Hints
Figure 5
Figure 6
Figure 8
Figure 7
Memory Mode
3.
The user can press the
Test Button
to take new measurements.
In the memory mode, mark or mark will not change.
1. The Memory Mode can be accessed either in ear mode or object
mode:
When the thermometer has been turned on and followed by
Figure 2/4 or finished testing, press the
. The
letter MEM will appear at the
of the display.
(See Figure )
Memory
Button
top right corner
7
2. The thermometer will automatically memorize the last 30
temperature readings.
Each time the
Button
is pressed, the screen displays past readings that correspond
with a number 1-30. The number 1 reflects the most recent
reading, while the number 30 reveals the oldest reading stored
in memory.(See Figure )
Each memory also records the
measurement date/time/mode icon.
Memory
8
After measurement:
1. Power off: Device will automatically shut off if left idle for
more than 1 minute to extend battery life.
2. Clean the probe after each use to ensure an accurate reading and
avoid cross contamination.
(See the section of Care and Cleaning for details.)
BackLight:
In Ear mode:
1. The display will be lighted GREEN for 3 seconds when the unit
is ready for measurement and a measurement is completed with
a reading less than 37.3
(99.1 ).
2. The display will be lighted YELLOW for 3 seconds when a
measurement is completed with a reading less than
3. The display will be lighted RED for 3 seconds when a measurement
is completed with a reading equal to or higher than
In Object mode:
The display will only be lighted GREEN for 3 seconds when the
unit is ready for measurement and a measurement is completed.
℃
℉
38.0
℃
(100.4 ).
38.0
℃
(100.4
℉
).
℉
Illustration For Use