background image

Calibration

The thermometer is initially calibrated at the time of manufacture. 
If the thermometer is used according to the use instruction, periodic 
readjustment is not required. However, We recommends checking 
calibration every two years or whenever clinical accuracy of the 
thermometer is in question. Please send the complete device to the 
dealers or manufacturer. 

The above recommendations do not supersede the legal requirements. 
The user must always comply with legal requirements for the control 
of the measurement, functionality, and accuracy of the device which 
are required by the scope of relevant laws, directives or ordinances 
where the device is used.

A clinical summary and procedures for checking calibration are 
available upon request.(Turn on the thermometer and press the 

Button

 long time until entering into calibrate mode, software

 

version

will be displayed.)

Test 

Troubleshooting

In Ear mode:
Temperature taken is 
higher than 43.0 

(109.4

).

In Object mode:
Temperature taken is 
higher than 100 

(212

).

Problem

Solution

Read Temperature 
Taking Hints thoroughly ,
then make sure the probe 
cover and lens filter are 
clean, then  take a new 
temperature measurement. 

The  thermometer could 
not

 

work due to low 

battery.

Replace two new 
alkaline batteries 
size 

.

 

AAA

Error message              

Read Temperature Taking 
Hints Thoroughly, then 
check the integrity of the 
probe cover and take a new 
temperature measurement. 

24

In Ear mode:
Temperature taken is 
lower than 32.0 

(89.6

).

In Object mode:
Temperature taken is 
lower than 0 

(32

).

23

26

25

Symbol Explanation

Troubleshooting

Problem

Solution

The ambient temperature
 is not within the range 
between 10

 and 40

(50

~

104

).

Place the 
 in a room for at least 30
 minutes at room 
temperature between 
10

 and 40

(50

~

104

)

 thermometer

The  thermometer is 
placed incorrectly
or unsteady. 

The  thermometer
 showing a rapid ambient 
temperature change.  

The  thermometer is not 
functioning properly.

Read Illustration For 
Use thoroughly and 
take a new temperature
 measurement.
 

Allow the thermometer 
to rest in a room for at 
least 30 minutes at 
room temperature: 
between 10

 and 40

(50

~

104

)

 Unload the battery, 

wait for 1 minute and 
repower it. If the 
message reappears, 
contact the retailer 
for service.
 

Error message              

Take a measurement 
until      or     appears 
on the display.

Measurement before 
thermometer is ready

Caution 

95%

15%

131

o

F

-13

o

F

Manufacturer

Manufacturing

 

Date

 

 Direct Current 

Batch Code

Storage and Transportation Temperature 
Limit: -13

~131

(-25

~55

℃)

TYPE BF APPLIED PART

Refer

 

to

 

instruction

 

manual

/

booklet

General

 

symbol

 

for

 

recovery

/

recyclable

Disposal

 

of

 

this

 

product

 

and

 

used

 

batteries

 

should

 

be

 

carried

 

out

 

in

 

accordance

 

with

 

the

 

national

 

regulations

 

for

 

the

 

disposal

 

of

 

electronic

 

products

.

106kPa

70kPa

Atmospheric pressure limitation

Storage and Transportation  
15%~95%RH

Humidity limitation: 

The first num.2:Protected against solid foreign objects of 12,5 
mm      and greater. 
The second num.2:Protection against vertically falling water 
drops when ENCLOSUREtilted up to 15

°

.

IP22 

Содержание DET-5000

Страница 1: ... eardrum of human It converts the measured heat into a temperature reading displayed on the LCD When properly used it will quickly assess your temperature in an accurate manner 5000 Document No Article No Version Z Date of Issue JDET 2504 002 001 2020 08 Telephone 86 571 81957767 Fax 86 571 81957750 JOYTECH Healthcare Co Ltd No 365 Wuzhou Road Yuhang Economic Development Zone Hangzhou Zhejiang 311...

Страница 2: ...surements Each memory also records the measurement results date time mode When the data exceeds 30 sets the coverage starts from the earliest data Beep Hint When the testing is done it will sound one short beep The real time clock will be recorded with the memory function and help you to recognize each measurement result Please see the Real time clock setting section to learn how to setup the time...

Страница 3: ...her an AM PM 12 hour or a 24 00 24 hour format Press and release Test button to select the format With the preferred time format on the display press Setting Button the Hour figure is flashing automatically Set the hour Press and release the Test to advance one hour until the correct hour appears After the hour is set press Setting Button the Minute figure is flashing automatically Button Vital Or...

Страница 4: ...e and object mode 1 Press the Test the display is activated to show all segments Button After self checking Figure 2 appears on the display screen with beeps so you can start a new measurement If error messages appears it means the thermometer is not ready for measurement 2 Insert the probe along the ear canal as deep as possible into the ear as depicted in Figure 3 Then press the Test BUTTON once...

Страница 5: ...ry Mode 3 The user can press the Test Button to take new measurements In the memory mode mark or mark will not change 1 The Memory Mode can be accessed either in ear mode or object mode When the thermometer has been turned on and followed by Figure 2 4 or finished testing press the The letter MEM will appear at the of the display See Figure Memory Button top right corner 7 2 The thermometer will a...

Страница 6: ... withstands drop of 3 feet Expected service life Three years Ear mode 32 0 43 0 89 6 109 4 Object mode 0 100 32 212 Ear mode 0 2 0 4 during 35 5 42 0 95 9 107 6 at 15 35 59 0 95 0 operating temperature range 0 3 0 5 for other measuring and operating temperature range Object mode 4 or 2 4 whichever is greater Battery Replacement 1 Replace battery when appears in the of LCD display See Figure top ri...

Страница 7: ...easurement 24 In Ear mode Temperature taken is lower than 32 0 89 6 In Object mode Temperature taken is lower than 0 32 23 26 25 Symbol Explanation Troubleshooting Problem Solution The ambient temperature is not within the range between 10 and 40 50 104 Place the in a room for at least 30 minutes at room temperature between 10 and 40 50 104 thermometer The thermometer is placed incorrectly or unst...

Страница 8: ...portable exposure condition without restriction FCC Information Warranty Thermometer is warranted by manufacture to be free from defects in material and workmanship under normal use and service for a period of one year from the date of delivery to the first user who purchases the instrument This warranty does not cover batteries damage to the probe window or damage to the instrument caused by misu...

Страница 9: ...the conditions described in the table below The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use Portable and mobile HF communications equipment can affect the device Use of the unit in conjunction with non approved accessories can affect the device negatively and alter the electromagneti...

Страница 10: ...4 MHz 7 MHz to 7 3 MHz 10 1 MHz to 10 15 MHz 14 MHz to 14 2 MHz 18 07 MHz to 18 17 MHz 21 0MHz to 21 4MHz 24 89 MHz to 24 99 MHz 28 0 MHz to 29 7 MHz and 50 0 MHz to 54 0 MHZ b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 7 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause inte...

Отзывы: