Calibration
The thermometer is initially calibrated at the time of manufacture.
If the thermometer is used according to the use instruction, periodic
readjustment is not required. However, We recommends checking
calibration every two years or whenever clinical accuracy of the
thermometer is in question. Please send the complete device to the
dealers or manufacturer.
The above recommendations do not supersede the legal requirements.
The user must always comply with legal requirements for the control
of the measurement, functionality, and accuracy of the device which
are required by the scope of relevant laws, directives or ordinances
where the device is used.
A clinical summary and procedures for checking calibration are
available upon request.(Turn on the thermometer and press the
Button
long time until entering into calibrate mode, software
version
will be displayed.)
Test
Troubleshooting
In Ear mode:
Temperature taken is
higher than 43.0
℃
(109.4
℉
).
In Object mode:
Temperature taken is
higher than 100
℃
(212
℉
).
Problem
Solution
Read Temperature
Taking Hints thoroughly ,
then make sure the probe
cover and lens filter are
clean, then take a new
temperature measurement.
The thermometer could
not
work due to low
battery.
Replace two new
alkaline batteries
size
.
AAA
Error message
Read Temperature Taking
Hints Thoroughly, then
check the integrity of the
probe cover and take a new
temperature measurement.
24
In Ear mode:
Temperature taken is
lower than 32.0
℃
(89.6
℉
).
In Object mode:
Temperature taken is
lower than 0
℃
(32
℉
).
23
26
25
Symbol Explanation
Troubleshooting
Problem
Solution
The ambient temperature
is not within the range
between 10
℃
and 40
℃
(50
℉
~
104
℉
).
Place the
in a room for at least 30
minutes at room
temperature between
10
℃
and 40
℃
(50
℉
~
104
℉
)
thermometer
The thermometer is
placed incorrectly
or unsteady.
The thermometer
showing a rapid ambient
temperature change.
The thermometer is not
functioning properly.
Read Illustration For
Use thoroughly and
take a new temperature
measurement.
Allow the thermometer
to rest in a room for at
least 30 minutes at
room temperature:
between 10
℃
and 40
℃
(50
℉
~
104
℉
)
Unload the battery,
wait for 1 minute and
repower it. If the
message reappears,
contact the retailer
for service.
Error message
Take a measurement
until or appears
on the display.
Measurement before
thermometer is ready
Caution
95%
15%
131
o
F
-13
o
F
Manufacturer
Manufacturing
Date
Direct Current
Batch Code
Storage and Transportation Temperature
Limit: -13
℉
~131
℉
(-25
℃
~55
℃)
TYPE BF APPLIED PART
Refer
to
instruction
manual
/
booklet
General
symbol
for
recovery
/
recyclable
Disposal
of
this
product
and
used
batteries
should
be
carried
out
in
accordance
with
the
national
regulations
for
the
disposal
of
electronic
products
.
106kPa
70kPa
Atmospheric pressure limitation
Storage and Transportation
15%~95%RH
Humidity limitation:
The first num.2:Protected against solid foreign objects of 12,5
mm and greater.
The second num.2:Protection against vertically falling water
drops when ENCLOSUREtilted up to 15
°
.
IP22