Illustration For Use

Illustration For Use

14

Illustration For Use

Figure 3

X

3.  The thermometer is ready for testing again once an ear mark 

     displays on the LCD.  A waiting period of 20 seconds between 

     testing is recommended to avoid excessive cooling of the skin.

4.  The thermometer will shut off automatically after 60 seconds 

     of inactivity.  To prolong battery life, press and hold 

 

to turn the unit off.

Memory

 

     

Button

13

12

11

To measure object temperature:

1. Aim the thermometer at the center of the object you want to 

    measure with a distance less than 5cm. 

2. Press the 

Test Button 

to turn on the thermometer, 

you can take

    the object temperature after hearing two beep sounds.(see figure 4)

    

    Remarks:

    This mode shows the actual, unadjusted surface temperatures, 

    which is different from the body temperature. 

then press and 

    hold 

Setting Button

 to switch to object mode, 

If error messages appears, it means the thermometer is not ready for 

    measurement. 

Figure 4

How to change the ear mode and object mode:

You can press and hold 

Setting Button 

to switch the mode between 

ear mode and object mode.

1.  Press the 

Test

 

,  the display is activated to show all 

     segments. 

Button

After self-checking Figure 2 appears on the display 

     screen with beeps, so you can start a new measurement. If error 

     messages appears, it means the thermometer is not ready for 

     measurement.

2.  Insert the probe along the ear canal as deep as possible into the 

     ear as depicted in Figure 3.  Then press the 

Test BUTTON

 once

     to start the measurement. A long beep will sound approximately 

     1 second after the 

Test BUTTON

 is activated. This signals the 

     end of the process as results are displayed on the LCD screen.

To measure ear temperature:

Figure 2

Real Time Clock Setting

 

Set the year

Press and release the 

Test 

to advance

one year until the correct year appears. 
After the year is set, press 

Setting Button

, the Month

figure will appear

.

 

Button  

 

Set the month

Press and release the 

Test 

 to advance 

one month until the correct month appears.
After the month is set, press 

Setting Button

, the Date

figure is flashing automatically

.

  

Button 

 

Set the date

Press and release the 

Test 

 to advance 

one day until the correct month appears.
After the day is set, press  

Setting Button 

to exit the

setting mode.

Button

How to change temperature unit :

You can press 

Setting Button 

to change the unit between 

℃ 

and

 ℉.

Содержание DET-5000

Страница 1: ... eardrum of human It converts the measured heat into a temperature reading displayed on the LCD When properly used it will quickly assess your temperature in an accurate manner 5000 Document No Article No Version Z Date of Issue JDET 2504 002 001 2020 08 Telephone 86 571 81957767 Fax 86 571 81957750 JOYTECH Healthcare Co Ltd No 365 Wuzhou Road Yuhang Economic Development Zone Hangzhou Zhejiang 311...

Страница 2: ...surements Each memory also records the measurement results date time mode When the data exceeds 30 sets the coverage starts from the earliest data Beep Hint When the testing is done it will sound one short beep The real time clock will be recorded with the memory function and help you to recognize each measurement result Please see the Real time clock setting section to learn how to setup the time...

Страница 3: ...her an AM PM 12 hour or a 24 00 24 hour format Press and release Test button to select the format With the preferred time format on the display press Setting Button the Hour figure is flashing automatically Set the hour Press and release the Test to advance one hour until the correct hour appears After the hour is set press Setting Button the Minute figure is flashing automatically Button Vital Or...

Страница 4: ...e and object mode 1 Press the Test the display is activated to show all segments Button After self checking Figure 2 appears on the display screen with beeps so you can start a new measurement If error messages appears it means the thermometer is not ready for measurement 2 Insert the probe along the ear canal as deep as possible into the ear as depicted in Figure 3 Then press the Test BUTTON once...

Страница 5: ...ry Mode 3 The user can press the Test Button to take new measurements In the memory mode mark or mark will not change 1 The Memory Mode can be accessed either in ear mode or object mode When the thermometer has been turned on and followed by Figure 2 4 or finished testing press the The letter MEM will appear at the of the display See Figure Memory Button top right corner 7 2 The thermometer will a...

Страница 6: ... withstands drop of 3 feet Expected service life Three years Ear mode 32 0 43 0 89 6 109 4 Object mode 0 100 32 212 Ear mode 0 2 0 4 during 35 5 42 0 95 9 107 6 at 15 35 59 0 95 0 operating temperature range 0 3 0 5 for other measuring and operating temperature range Object mode 4 or 2 4 whichever is greater Battery Replacement 1 Replace battery when appears in the of LCD display See Figure top ri...

Страница 7: ...easurement 24 In Ear mode Temperature taken is lower than 32 0 89 6 In Object mode Temperature taken is lower than 0 32 23 26 25 Symbol Explanation Troubleshooting Problem Solution The ambient temperature is not within the range between 10 and 40 50 104 Place the in a room for at least 30 minutes at room temperature between 10 and 40 50 104 thermometer The thermometer is placed incorrectly or unst...

Страница 8: ...portable exposure condition without restriction FCC Information Warranty Thermometer is warranted by manufacture to be free from defects in material and workmanship under normal use and service for a period of one year from the date of delivery to the first user who purchases the instrument This warranty does not cover batteries damage to the probe window or damage to the instrument caused by misu...

Страница 9: ...the conditions described in the table below The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use Portable and mobile HF communications equipment can affect the device Use of the unit in conjunction with non approved accessories can affect the device negatively and alter the electromagneti...

Страница 10: ...4 MHz 7 MHz to 7 3 MHz 10 1 MHz to 10 15 MHz 14 MHz to 14 2 MHz 18 07 MHz to 18 17 MHz 21 0MHz to 21 4MHz 24 89 MHz to 24 99 MHz 28 0 MHz to 29 7 MHz and 50 0 MHz to 54 0 MHZ b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 7 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause inte...

Отзывы: