SECTION 4 | TECHNICAL REFERENCES

4.3 | CONTROL BOARD DIAGNOSTIC

LED Code

Possible Cause

Fast Flash

Control Module is not calibrated,  Do not operate unit.

Steady

Unit has just been powered on.  You may need to wait for initialization, then re-select function.
Ready to operate, things should be working normally.
A function is selected but the enable trigger is not squeezed.

1-1

The control module is not calibrated.  Do not use this unit.

2-1

The key switch selector switch indicate the mode in which the TS100 must operate. If neither 
input is active, or if both are active together, the TS100 does not know how to function. 
Check key switch and wiring to P15-1 and P12-1.

2-2

A safety feature is locking functions or a switch has failed.
Check that platform is not overloaded, operating on a level surface, and pothole guards deploy/
Check that joystick is neutral when powered on.
Check that joystick trigger is not closed for too long without selecting a function.
Check for failed joystick, selector switches, and up/down switches.

3-x

There is a problem with the drive contactor or valve wiring, or with the motor power wiring; 
disconnect connector P9 to see if the problem is caused by drive contactor or valve wiring (if the 
fault clears, check for an illegal B+ supply in to P9) 
Check motor power wiring; with the drive contactor open the B+ power terminals should be at 
10V-15V (significantly lower than B+) If the LED is steady at power-on, and the fault (3-5) occurs 
after a delay when attempting to drive or lift, the motor may be stalled and causing an overload of 
the TS100 or there is a power wiring error like connecting the B+ cable to a motor stud

3-2

Check P9 wiring.  One or more signals showing outputs when all should be off.

3-3

Check B+ stud connections on controller.  Voltage is too high.

3-4

There is voltage on safe pre-valve supply when there should not be.
Controller may need to be replaced.

3-5

The drive brake current is too high.
Motor overload.  Check for a siezed motor or for power wiring to motors.

4-x

There is a problem with battery supply, the height and/or pressure sensors, the supply to them, or
the temperature sensor inside the TS100
Check battery supply to EMS inputs P15-1 or P12-1 (relative to the B- stud); the battery supply
should be between 15V and 32V
Check the output from height sensor (P12-12)
If the TS100 heatsink is very hot then perhaps the controller has temporarily shut down – if so,
platform lowering is still allowed; wait for the controller to cool down

4-2

Functions Locked: Board is overheated.  Check pump, drive motor wiring.
Problem with controller internal voltage. Controller may need to be replaced.

When using the LED for diagnosis, note that a DUAL FLASH code is 
indicated. The LED will flash on/off a certain number of times, pause 
off for a short delay, then flash on/off a second certain number of 
times, followed by a much longer pause off. The sequence will then 
repeat. 

EXAMPLE: The LED flash code 3-2 will look like:  on/off/on/off/on/
off-short-delay/on/off/on/off-long-delay/repeat         

MAINTENANCE & TROUBLESHOOTING
HB-1030CE/HB-1430CE

SUPO-714

REV C

20

SECTION 4 | TECHNICAL REFERENCES

4-3

Problem with controller internal voltage. Controller may need to be replaced.

4-4

Battery supply is too low or too high.  
Make sure batteries afe fully charged.  
Do not operate while charging.

4-5

Joystick signal problem.
Wiring problem-check for short circuits, misconnection, check P15-12 connection.

6-x

There is a problem with the height measurements, load measurements, or the elevation switch 
disagrees with the height sensor.
Check that the output from height sensor (P12-12) is in range (between 0.5V and 4.5V)

6-1

Problem with angle sensor or its connections

6-2

Problem with the pressure sensor

6-3

Problem with elevation switch or its connections

6-6

Problem with the pressure sensor 

7-x

There is a problem with the power wiring – the voltage on the B+ stud is too low 
Check for a short-circuit to the B+ stud

7-1

Motor A current too high.

7-2

Motor A current too low.

7-3

Motor B current too high.

7-4

Motor B current too low.

7-5

Check drive connections at both drives--short or multiple wiring faults.

7-7

Check B+ stud connections on controller.  Voltage is too low.

MAINTENANCE & TROUBLESHOOTING
HB-1030CE/HB-1430CE

SUPO-714

REV C

21

Содержание Hy-Brid Lifts HB 1030CE

Страница 1: ...MAINTENANCE TROUBLESHOOTING MANUAL SELF PROPELLED AERIAL WORK PLATFORM SUPO 714 REV C HB 1030CE HB 1430CE SERIES III BY CUSTOM EQUIPMENT LLC...

Страница 2: ...and operating procedures essential to properly and safely operate the Custom Equipment Hy Brid Lift Users must read and understand all information in the Safety and Operations Manual before operation...

Страница 3: ...RENCES 16 4 1 HYDRAULIC SCHEMATIC HS HBCE 16 4 2 ELECTRICAL SCHEMATIC WS 112 20 301 51 18 4 3 CONTROL BOARD DIAGNOSTIC 20 SECTION 5 WIRING DIAGRAMS 22 5 1 WIRING DIAGRAM PART NO WD 112 20 301 51 22 5...

Страница 4: ...think might be unsafe stop the platform and request further information from quali ed sources before proceeding MAINTENANCE INFORMATION IS FOR USE BY TRAINED PERSONNEL ONLY NEVER REACH BETWEEN SCISSO...

Страница 5: ...tery EXAMPLE A battery that is 100 charged at 80 F 27 C drops to 65 at 32 F 0 C At 0 F 18 C this battery will drop to 40 efficiency NEVER ADD ACID TO BATTERY Battery Maintenance WET CELL BATTERIES The...

Страница 6: ...E HB 1430CE SUPO 714 REV C 10 2 3 LUBRICATION Item Speci cation Frequency of Lubrication Wheels Te on Spray Quarterly Optional 2 4 COMPONENTS REQUIRING ADJUSTMENT Under normal use no components should...

Страница 7: ...tform Controls Check all switches and push buttons for proper operation Emergency Stop Stops all movement For Actuator Steered models Enable Switch Does not elevate unless enable is pressed For Counte...

Страница 8: ...d models Tie rods secure All Functions Srive Elevate Steer Operational see Pre Start Inspection for details Wheels Pothole guards deploy when platform elevated Snap Rings Secure Emergency Stop Breaks...

Страница 9: ...HNICAL REFERENCES 4 1 HYDRAULIC SCHEMATIC HS HBCE MAINTENANCE TROUBLESHOOTING HB 1030CE HB 1430CE SUPO 714 REV C 16 SECTION 4 TECHNICAL REFERENCES MAINTENANCE TROUBLESHOOTING HB 1030CE HB 1430CE SUPO...

Страница 10: ...0 0 0 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 5 7 6 12 12 6 0 5 5 5 5 12 5 9 1 3ODWIRUP WO 6HOHFW DVH WO 6HOHFW DVH LIW 6LJQDO DVH RZHU 6LJQDO UDNH 2XWSXW 6 9 6 9 6 LS...

Страница 11: ...roblem with battery supply the height and or pressure sensors the supply to them or the temperature sensor inside the TS100 Check battery supply to EMS inputs P15 1 or P12 1 relative to the B stud the...

Страница 12: ...3 7 MAINTENANCE TROUBLESHOOTING HB 1030CE HB 1430CE SUPO 714 REV C 22 SECTION 5 WIRING DIAGRAMS 0 526 7 5 0 5 9 5 0 526 7 5 9 0 77 5 77 5 5 12 1 3 0 7 6 7 6 1 0 5 21 36 36 1 6 1625 1 6 1625 2 5 0 360...

Страница 13: ...51 MAINTENANCE TROUBLESHOOTING HB 1030CE HB 1430CE SUPO 714 REV C 24 SECTION 5 WIRING DIAGRAMS 7 5 12 12 6 7 7 5 5 5 5 5 2 1 2 WR WR 6 6 WR 68 72 78 725 5 7 67 5 0 7 6 7 7 67 5 0 7 6 7 WR WR WR WR WR...

Страница 14: ...E SUPO 714 REV C 26 SECTION 5 WIRING DIAGRAMS 7 5 7 1 7 5277 7 5277 5 9 0 1 211 7 21 2 67 211 7 21 2 67 3 7 211 725 5217 9 6 25 1 5 1 XVWRP TXLSPHQW QF 5LFKILHOG 3KRQH 7 835 7 6 2 127 6 5 1 7 7 21 7 7...

Страница 15: ...blem Yes Reference Revision A Are all power supply connections and switches functioning properly isually inspect wire harness from lower controls to upper Is there damage y switch on Reset E Stop butt...

Страница 16: ...ting If a part has been identified as needing replacement see the Parts View To identify part number to order If any wiring or components have been altered from the original manufacture problems may n...

Страница 17: ...nections at actuator Pull test connections if failure crimp on new connector s Consider mode limit switch wiring or switch failure red and or black wires Continuity test on switch voltage test on red...

Страница 18: ...ilable Yes Replace Pump Assembly Does pump operate Try replacing up valve No Is there a different problem something is leaking unusually noisy etc Replace pump assembly May be worn or defective Troubl...

Страница 19: ...cylinder or damaged seal in cylinder Run unit up and then check for oil flow out of return line Bad cylinder seal if oil is flowing from return line Yes Faulty check valve in pump Listen carefully ne...

Страница 20: ...430CE SUPO 714 REV C 38 Description Part Number Notes ALARM CONTINUOUS ELEC 635 4 BOARD DRIVE LIFT CTL HB MID 112 21 311 51 BUTTON PUSH PULL RED E STOP ELEC 071 KIT CHARGER 24V ELEC 747 CORD NEMA 515...

Страница 21: ...s or parts components or assemblies thereof that are replaced under this Limited Warranty shall become the property of the Company The Company reserves the right to review Buyer s maintenance and oper...

Страница 22: ...NOTES MAINTENANCE TROUBLESHOOTING HB 1030CE HB 1430CE SUPO 714 REV C 42 NOTES MAINTENANCE TROUBLESHOOTING HB 1030CE HB 1430CE SUPO 714 REV C 43...

Страница 23: ...647 Highway 175 Richfield WI 53076 U S A Tel 1 262 644 1300 Fax 1 262 644 1320 www hybridlifts com Service Customequipmentlifts com Hy Brid Lifts is a trademark of Custom Equipment LLC These machines...

Отзывы: