C
HAPTER
3: C
ONFIGURING
THE
AXN
SYSTEM
35
Maximum Balance defines the required headroom to be applied on top of the user-specified range for each
channel. The setup software sets instrumentation amplifier gain to allow for the required headroom. Ensure you do
not define a maximum balance that pushes the channel outside the range of the instrumentation amplifier. The
Maximum Balance unit is input mode dependent; it is expressed in volts for differential ended, single ended, full-
bridge, half-bridge, AC coupled, and ICP modes. For Thermocouple and RTD modes, the range is in Celsius.
Balance Target defines the value the balance process attempts to adjust the channel counts to match. The
Balance Target unit is input mode dependent, it is expressed in volts for differential ended, single ended, full-
bridge, half-bridge, AC coupled and ICP modes. For Thermocouple and RTD modes the range is in Celsius.
Enabling/disabling shunt resistors
The AXN/ADC/401 supports shunt resistors per channel. These are 100 kΩ resistors that can be switches between
inputs and ground. These resistors are classically used to test bridge operations pre-flight, but may also
be used for ground reference when the input to the module channel is isolated from the module ground.
Configuring the AXN/ADC/401 for differential ended inputs
1.
Set the channel range to cover the range of the measurement you want.
N
OTE
:
This applies to all filter tap points for that channel.