C
HAPTER
3: C
ONFIGURING
THE
AXN
SYSTEM
33
For each filter tap point per channel, you can define the filter cutoff applied to each filter tap point independently.
Setting Excitation Mode
Excitation Mode can be independently set for each channel to either Current or Voltage excitation.
Excitation Amplitude is in volts for Voltage excitation and is limited to 0V to 10.2V. Excitation is unipolar on the
AXN/ADC/401. For example, to set 5V excitation current, set the Excitation Mode to Voltage and Excitation
Amplitude to 5.
The Excitation Amplitude is in amps for Current excitation and is limited to 0 mA to 32 mA. For example, to set
2 mA excitation current, set the Excitation Mode to Current and Excitation Amplitude to 0.002.