background image

4

GG-CO2

General Description

Installation

The GG-CO2 sensor is a +24 VDC, three-wire, 4/20 

mA sensor for carbon dioxide which utilizes proven 

infrared sensor technology for accurate and reliable 

leak detection.  The 0-3% (0-30,000ppm) detection 

range of the GG-CO2 provides real-time continuous 

monitoring of ambient carbon dioxide concentrations. 
0-1% and 0-5% ranges are also available. 
The GG-CO2 provides an industry standard linear 4/20 

mA output signal compatible with most gas detection 

systems and PLCs.  The output signal is not affected 

by drastic temperature and moisture variations that 

occur during washdown, defrost cycles, etc. 
The circuit board is sealed in potting compound, 

completely protecting sensitive electronic components 

and copper tracing from corrosion.  The specially 

vented chemical-resistant polycarbonate enclosure 

protects the sensor from accidental damage, weather 

and direct hose-hits from cleaning crews.

Locating the sensor

One of the most important considerations when 

installing GG-CO2 sensors is that they must be easily 

accessible for calibration and maintenance. 
For optimum personnel protection (representative 

concentration reading that an employee would be 

exposed to), mount the sensor at a height in the 

breathing zone of the employees.  It would typically 

be about five feet off the ground, which also allows 

easy access.  As a general rule of thumb, try to mount 

sensors within 30 feet of potential leak sources.

Содержание GG-CO2

Страница 1: ...Installation and Operation Manual GG CO2 CARBON DIOXIDE GAS SENSOR...

Страница 2: ...ly in the manner described in this manual If the equipment is used in a manner not specified by Calibration Technologies the protection provided by the equipment may be impaired This equipment should...

Страница 3: ...920 N Tradewinds Pkwy Columbia MO 65201 866 394 5861 Sales ctiengineering com www ctiengineering com General description 4 Installation 4 Locating the sensor 4 Installation guidelines 5 Wiring 6 Oper...

Страница 4: ...cycles etc The circuit board is sealed in potting compound completely protecting sensitive electronic components and copper tracing from corrosion The specially vented chemical resistant polycarbonat...

Страница 5: ...variable speed drives and radio repeaters Protect sensor from physical damage forklifts etc If mounting on a wall with studs the mounting screws should be screwed into the studs Mount sensor enclosur...

Страница 6: ...elded copper cable Do not pull sensor wiring with AC power cables This can cause electrical interference If cable runs cannot be made without a splice all splice connections should be soldered Ground...

Страница 7: ...asure sensor output in mA Always use mVDC or VDC voltmeter settings Zero Calibration After the unit is installed and has been powered up for a minimum of 2 hours the unit can be zero calibrated by the...

Страница 8: ...8 GG CO2 Figure 2 Sensor output and zero span adjustments Span and Zero adjustments 40 200 mV mVDC Black Red Calibration Gas Port...

Страница 9: ...ld be performed with certified calibration gas Calibration kits and replacement cylinders are available from Calibration Technologies All tests and calibrations must be logged Always disconnect power...

Страница 10: ...cale Repeatability 1 of full scale Wiring Connections 3 conductor shielded stranded 20 AWG cable General Cable C2525A or equivalent up to 1500 ft Terminal Block Plug Field Wiring 26 12 AWG torque 4 5...

Страница 11: ...ct by any person other than an authorized dealer or contractor or the installation of unapproved parts on the product The obligations set forth in this warranty are conditional on a proper storage ins...

Страница 12: ...ctiengineering com 866 394 5861 GG CO2 DOC1 0 20190604...

Отзывы: