GEM-II - Operation Manual
Rev. E | 2014.10
FUNCTION DESIRED
J6 LOWER
JUMPER BANK
J7 UPPER
JUMPER BANK
Resting
(Jumpers for the Upper Bank, J7 are disabled)
1
1
Setting Span Gas Value
2 (Internal Sensor)
2
Perform Zero (Null) Calibration
2 (Internal Sensor)
3
Perform Span Calibration
2 (Internal Sensor)
4
Adjust Low Gas Alarm Ascending Value
2 (Internal Sensor)
5
Adjust High Gas Alarm Ascending Value
2 (Internal Sensor)
6
Adjust Low Gas Alarm Descending Value
2 (Internal Sensor)
7
Adjust High Gas Alarm Descending Value
2 (Internal Sensor)
8
Enable / Disable Current Output
2 (Internal Sensor)
9
Setting Span Gas Value
3 (External Sensor)
2
Perform Zero (Null) Calibration
3 (External Sensor)
3
PERFORM SPAN CALIBRATION
3 (External Sensor)
4
Adjust Low Gas Alarm Ascending Value
3 (External Sensor)
5
Adjust High Gas Alarm Ascending Value
3 (External Sensor)
6
Adjust Low Gas Alarm Descending Value
3 (External Sensor)
7
Adjust High Gas Alarm Descending Value
3 (External Sensor)
8
Enable / Disable Current Output
3 (External Sensor)
9
NOTE:
Jumper - 4 location on the lower jumper bank (J6) is used to set custom time delays for
relays and internal audible alarm. These functions will be explained in another section of this
manual.