User Manual for Handheld Pulse Oximeter
20
Accuracy: ±2bpm or ±2% (whichever is greater)
Default alarm limits: High: 120bpm Low: 50bpm
Pulse Rate alarm limit range
:
30bpm~240bpm
E.
Perfusion Index Display
Range: 0~20%
F.
Update rate:
8 beats moving average for both SpO
2
and pulse rate readings
G.
Data Record
Record data once every 1/2/4/8seconds, up to 70–hour
records.
H.
The performance under low perfusion condition
The accuracy of SpO
2
and PR measurement still meet the
precision described above when the modulation amplitude
is as low as 0.5%.
I.
Resistance to interference of surrounding light:
The difference between the SpO
2
value measured in the
condition of indoor natural light and that of darkroom is less
than ±1%.
J.
The resistance to 50Hz /60Hz interference
SpO
2
and PR are precise which have been tested by
BIO-TEK pulse oximeter simulator.
K.
Dimensions:
145 mm (L) × 74 mm (W) × 29 mm (H)
Net Weight:
210g (including batteries)
L.
Classification
The type of protection against electric shock:
Содержание PC 66B