User Manual for Spot-Check Monitor
14
2. If NIBP and SpO
2
are measured simultaneously, then the NIBP measured result will be prior to be
displayed for 10 seconds.
CHAPTER 5 TECHNICAL SPECIFICATIONS
5.1 Blood Pressure Measurement
1. Blood pressure measuring range:
Systolic pressure: 60mmHg~240mmHg
Diastolic pressure: 30mmHg~180mmHg
2. Blood pressure measuring accuracy: mean difference≤5mmHg, standard deviation≤8mmHg
3. Pneumatic system error: ±3mmHg
4.
Pressure display resolution: 1mmHg
5.2 SpO
2
Measurement (Optional)
1. SpO
2
measuring range: 35%~100%
2. SpO
2
measuring accuracy: A
rms
is not greater than 2% for SpO
2
range from 70% to 100%
Note: A
rms
is defined as root-mean-square value of deviation according to ISO 9919