User Manual for Spot-Check Monitor
15
3. Resolution: 1%
4. Low limit alarm value for SpO
2
: 90%.
Note: when the measured SpO
2
≤ 90%, then the over-limit SpO
2
value blinks.
5.3 Pulse Rate Measurement (Optional)
1. Pulse Rate measuring range: 30bpm~240bpm
2 . Pulse rate measuring accuracy: ±2bpm or ±2% of reading, whichever is greater
3. Resolution: 1bpm
Note: when the measured PR ≥ 120bpm or PR ≤50bpm, then the over-limit PR value blinks.
5.4 Wireless Communication
1. Emit and receiving frequency: 2402MHz~ 2480MHz
2. The Channel number: 40
3. Bandwidth: 2 MHz
4. Modulation method: GFSK
Содержание PC-101