Use the Custom menu to enter the K-Factor and Offset for a custom sensor not
included in this meter.
K-Factor Calibration (
F a c t o r
)
The meter may be calibrated using the
K-Factor
function. Most flowmeter
manufacturers provide this information with the device. Enter the
K-Factor
(
Factor
) menu and select the decimal point with highest resolution possible and
program the k-factor value (
from sensor manual
). The meter will automatically
calculate the flow rate using the k-factor and the time base selected.
Select Rate and Total Units (
r a t e u n i t s
/
t o t a l u n i t s
)
Use the rate units and total units menus to select the appropriate rate and total
units for your application.
The units available for rate are:
1. Gallons per minute (
GPnm
)
2. Liters per second (
LPS
)
3. Liters per minute (
LPnm
)
The units available for total are:
1. Gallons (
GAL
)
2. Cubic meters (
nm 3
)
3. Cubic feet (
CuFt
)