DFM-6300 Series
Instruction Manual
4
INSTALLATION
There is no need to remove the meter from its case to complete the
installation, wiring, and setup of the meter for most applications.
Unpacking
Remove the meter from box. Inspect the packaging and contents for damage.
Report damages, if any, to the carrier.
If any part is missing or the meter malfunctions, please contact your supplier
or the factory for assistance.
Panel Mounting Instructions
•
Prepare a standard 1/8 DIN panel cutout – 3.622" x 1.772" (92 mm x 45 mm). Refer to
Figure 1 for more details.
•
Clearance: allow at least 6.0" (152 mm) behind the panel for wiring.
•
Panel thickness: 0.04" - 0.25" (1.0 mm - 6.4 mm).
Recommended minimum panel thickness to maintain Type 4X
rating: 0.06" (1.5 mm) steel panel, 0.16" (4.1 mm) plastic panel.
•
Remove the two mounting brackets provided with the meter (back-off the two screws so
that there is
¼
" (6.4 mm) or less through the bracket. Slide the bracket toward the front of
the case and remove).
•
Insert meter into the panel cutout.
•
Install mounting brackets and tighten the screws against the panel. To achieve a proper
seal, tighten the mounting bracket screws evenly until meter is snug to the panel along
its short side. DO NOT OVER TIGHTEN, as the rear of the panel may be damaged.
Note: See
Figure 10
on page 30 for
1/8 DIN Panel Cutout Templat
e
Figure 1: 1/8 DIN Panel Cutout and Mounting
Panel
Gasket
Mounting
Bracket
Mounting
Screw
Removable
Connectors
3.622" (92 mm)
1.772"
(45mm)
Panel Cutout
to DIN 43700
Square Corners to 0.060"
(1.5mm) Max Radius
A
B
Tolerances:
A: +0.032 (+0.8mm)
-0.000 (-0.0mm)
B: +0.024 (+0.6mm)
-0.000 (-0.0mm)