13
Crane
Electronics Ltd
The force in torque management
Crane Electronics Ltd
Watling Drive
Sketchley Meadows
Hinckley LE10 3EY
Tel: +44(0) 1455 25 14 88
www.crane-electronics.com
Peak / Track Measurement Mode
– Switch between Peak and Track measurement
modes.
Peak Mode
The peak mode displays and retains the maximum torque experienced by the wrench, as occurs when operating
the wrench in the tightening direction. The peak mode is recommended for rotary tools and wrenches.
Track Mode
The track mode displays the torque as it is being applied to the transducer. This mode is used primarily for
verifying the calibration of the WM10K unit.
Press the Menu button and then press
the left and right arrow keys to scroll
through the measurement units,
selecting either Nm or Ft.lbs.
Use this option to select the filter value.
Press the Menu button twice and using
the arrow keys, select the required
Filter. Selectable filter values in Hz (125,
250, 500, 1000, 2000, 4000 and 5000).
The filter is used to reduce mechanical
noise when used with impulse type
tools. Current ISO standards
recommend 500Hz for general use.
MEASUREMENT UNITS